Agilent Technologies InfiniiVision 3000 DSO-X 3054A Manuale di Servizio Pagina 832

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 970
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 831
832 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
34 Status Reporting
Table 66
Table 61
The status registers picture above shows how the different status reporting
data structures work together. To make it possible for any of the Standard
Event Status Register bits to generate a summary bit, the bits must be
enabled. These bits are enabled by using the *ESE common command to
set the corresponding bit in the Standard Event Status Enable Register.
To generate a service request (SRQ) interrupt to an external controller, at
least one bit in the Status Byte Register must be enabled. These bits are
enabled by using the *SRE common command to set the corresponding bit
in the Service Request Enable Register. These enabled bits can then set
RQS and MSS (bit 6) in the Status Byte Register.
Vedere la pagina 831
1 2 ... 827 828 829 830 831 832 833 834 835 836 837 ... 969 970

Commenti su questo manuale

Nessun commento