510 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
24 :SBUS<n> Commands
:SBUS<n>:LIN:STANdard
(see page 860)
Command Syntax
:SBUS<n>:LIN:STANdard <std>
<std> ::= {LIN13 | LIN20}
The :SBUS<n>:LIN:STANdard command sets the LIN standard in effect for
triggering and decoding to be LIN1.3 or LIN2.0.
Query Syntax
:SBUS<n>:LIN:STANdard?
The :SBUS<n>:LIN:STANdard? query returns the current LIN standard
setting.
Return Format
<std><NL>
<std> ::= {LIN13 | LIN20}
See Also "Introduction to :TRIGger Commands" on page 641
":TRIGger:MODE" on page 648
":SBUS<n>:LIN:SIGNal:DEFinition" on page 816
":SBUS<n>:LIN:SOURce" on page 509