:DISPlay Commands 13
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 261
:DISPlay:PERSistence
(see page 860)
Command Syntax
:DISPlay:PERSistence <value>
<value> ::= {MINimum | INFinite | <time>}
<time> ::= seconds in in NR3 format from 100E-3 to 60E0
The :DISPlay:PERSistence command specifies the persistence setting:
MINimum — indicates zero persistence.
INFinite — indicates infinite persistence.
<time> — for variable persistence, that is, you can specify how long
acquisitions remain on the screen.
Use the :DISPlay:CLEar command to erase points stored by persistence.
Query Syntax
:DISPlay:PERSistence?
The :DISPlay:PERSistence? query returns the specified persistence value.
Return Format
<value><NL>
<value> ::= {MIN | INF | <time>}
See Also "Introduction to :DISPlay Commands" on page 255
":DISPlay:CLEar" on page 257