Setting Up Generic and Advanced Signals Setting Up a Burst Signal
46 Agilent 81133A/81134A Pulse Generator User’s Guide, September 2008
Setting Up a Burst Signal
Task Set up a burst signal consisting of a 32- bit data pattern repeated
twice at a period of 500 ps with 50 ps delay and 100 ps pulse
width.
The data pattern is: 10100000000000000000000000000000
The amplitude is 2.0 V and the offset is 0 V.
Start the output at the rising edge of an external signal with 1 V
threshold applied at the start input.
Using the Graphical User Interface
Instrument Settings To set the instrument parameters:
1 Disable the outputs.
2 In the instrument panel, choose the Burst mode.
3 Set the burst repetition to 2.
4 Set the Period to 500 ps.
Arming Source
Triggered by:
External Signal
-1 V
Generated
Signal
1 V
Delay = 50 ps
Width = 100 ps
Period = 500 ps