User’s and Service GuideAgilent Technologies 85032B/E 50Ω Type-N Calibration KitsAgilent Part Number: 85032-90020Printed in USA Print Date: August 201
1-6 85032B/E General InformationClarifying the Terminology of a Connector InterfaceClarifying the Terminology of a Connector InterfaceIn this document
85032B/E 1-7General InformationWhen to CalibrateWhen to CalibrateA network analyzer calibration remains valid as long as the changes in the systematic
1-8 85032B/E General InformationWhen to Calibrate
85032B/E 2-12 Specifications
2-2 85032B/E SpecificationsEnvironmental RequirementsEnvironmental Requirements Temperature—What to Watch Out ForChanges in temperature can affect ele
85032B/E 2-3SpecificationsMechanical CharacteristicsMechanical CharacteristicsMechanical characteristics such as center conductor protrusion and pin d
2-4 85032B/E SpecificationsMechanical CharacteristicsNOTEThe gages for measuring type-N connectors compensate for the designed offset of 5.26 mm (0.20
85032B/E 2-5SpecificationsElectrical SpecificationsElectrical SpecificationsThe electrical specifications in Table 2-3 apply to the devices in your ca
2-6 85032B/E SpecificationsElectrical Specifications
85032B/E 3-13 Use, Maintenance, and Care of the Devices
ii 85032B/E Documentation WarrantyTHE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED "AS IS," AND IS SUBJECT TO BEING CHANGED, WITHOUT NOT
3-2 85032B/E Use, Maintenance, and Care of the DevicesElectrostatic DischargeElectrostatic DischargeProtection against ESD (electrostatic discharge) i
85032B/E 3-3Use, Maintenance, and Care of the DevicesVisual InspectionVisual InspectionVisual inspection and, if necessary, cleaning should be done ev
3-4 85032B/E Use, Maintenance, and Care of the DevicesCleaning ConnectorsInspect Female ConnectorsPay special attention to the contact fingers in the
85032B/E 3-5Use, Maintenance, and Care of the DevicesCleaning ConnectorsUse a lint-free swab or cleaning cloth moistened with isopropyl alcohol to rem
3-6 85032B/E Use, Maintenance, and Care of the DevicesGaging ConnectorsGaging ConnectorsThe gages available from Agilent Technologies are intended for
85032B/E 3-7Use, Maintenance, and Care of the DevicesGaging ConnectorsWhen to Gage ConnectorsGage a connector at the following times:• Prior to using
3-8 85032B/E Use, Maintenance, and Care of the DevicesGaging ConnectorsGaging ProceduresGaging Male Type-N ConnectorsNOTEAlways hold a connector gage
85032B/E 3-9Use, Maintenance, and Care of the DevicesGaging ConnectorsFigure 3-3 Gaging Male Type-N Connectors
3-10 85032B/E Use, Maintenance, and Care of the DevicesGaging ConnectorsGaging Female Type-N ConnectorsNOTEAlways hold a connector gage by the gage ba
85032B/E 3-11Use, Maintenance, and Care of the DevicesGaging ConnectorsFigure 3-4 Gaging Female Type-N Connectors
Contents85032B/E Contents-iii1 General InformationCalibration Kit Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3-12 85032B/E Use, Maintenance, and Care of the DevicesConnectionsConnectionsGood connections require a skilled operator. The most common cause of mea
85032B/E 3-13Use, Maintenance, and Care of the DevicesConnectionsUsing a torque wrench guarantees that the connection is not too tight, preventing pos
3-14 85032B/E Use, Maintenance, and Care of the DevicesConnections5. Apply downward force perpendicular to the wrench handle. See Figure 3-6. This app
85032B/E 3-15Use, Maintenance, and Care of the DevicesHandling and StorageHow to Separate a ConnectionTo avoid lateral (bending) force on the connecto
3-16 85032B/E Use, Maintenance, and Care of the DevicesHandling and Storage
85032B/E 4-14 Performance Verification
4-2 85032B/E Performance VerificationIntroductionIntroductionThe performance of your calibration kit can only be verified by returning the kit to Agil
85032B/E 4-3Performance VerificationRecertificationRecertificationThe following will be provided with a recertified kit:• a new calibration sticker af
4-4 85032B/E Performance VerificationRecertification
85032B/E 5-15 Troubleshooting
Contents-iv 85032B/EContentsHandling and Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5-2 85032B/E TroubleshootingTroubleshooting ProcessTroubleshooting ProcessIf you suspect a bad calibration, or if your network analyzer does not pass
85032B/E 5-3TroubleshootingWhere to Look for More InformationWhere to Look for More InformationThis manual contains limited information about network
5-4 85032B/E TroubleshootingContacting AgilentContacting AgilentAssistance with test and measurements needs and information on finding a local Agilent
85032B/E 6-16 Replaceable Parts
6-2 85032B/EReplaceable PartsIntroductionIntroductionTable 6-1 lists the replacement part numbers for items in the 85032B calibration kit.Table 6-2 l
85032B/E 6- 3Replaceable PartsIntroductionFigure 6-1 Replaceable Parts for the 85032B Calibration Kit
6-4 85032B/EReplaceable PartsIntroductionFigure 6-2 Replaceable Parts for the 85032E Calibration KitTable 6-2 Replaceable Parts for the 85032E Ca
85032B/E 6- 5Replaceable PartsIntroductionTable 6-3 Replaceable Parts—Items Not Included in the Calibration Kit Description Qty Agilent Par
6-6 85032B/EReplaceable PartsIntroduction
85032B/E A-1A Standard Definitions
85032B/E 1-11 General Information
A-2 85032B/E Standard DefinitionsClass Assignments and Standard Definitions Values are Available on the WebClass Assignments and Standard Definitions
Index85032B/E Index-1Numerics8510 network analyzer, 1-3Aadapters, 1-2part numbers, 1-5Agilent Technologiesapplication note, A-2contacting, 5-3, 5-4alc
Index-2 85032B/EIndexFfemale open, 3-14connecting, 3-14disconnecting, 3-14frequencyspecifications, 2-5frequency of calibration, 1-7frequency range, 2-
Index85032B/E Index-3supplemental electrical characteristics, 2-5suppliescleaning, 1-4Ttagservice, 1-4, 4-3, 5-3temperaturecalibration, 2-2device, 2-2
Index-4 85032B/EIndex
1-2 85032B/E General InformationCalibration Kit OverviewCalibration Kit OverviewThe Agilent 85032B and 85032E type-N calibration kits are used to cali
85032B/E 1-3General InformationCalibration Kit Overview The calibration definitions can be:• resident within the analyzer• entered from the front pane
1-4 85032B/E General InformationEquipment Required but Not SuppliedEquipment Required but Not SuppliedSome items are required or recommended for succe
85032B/E 1-5General InformationRecording the Device Serial NumbersRecording the Device Serial NumbersIn addition to the kit serial number, the devices