A Agilent 3000 Series OscilloscopesUser’s and Service Guide

Pagina 2

10 Agilent 3000 Series Oscilloscopes User’s Guide Contents6 Oscilloscope Utility Settings 111Performing Mask Tests 112To enable/disable mask tests 112

Pagina 3

4 Making Measurements100 Agilent 3000 Series Oscilloscopes User’s GuideIf both A and B cursors are used, these values are also displayed:• DeltaX — d

Pagina 4 - In This Book

A101Agilent 3000 Series OscilloscopesUser’s and Service Guide 5Saving, Recalling, and Printing DataSaving and Recalling Waveforms and Setups 102Savi

Pagina 5

5 Saving, Recalling, and Printing Data102 Agilent 3000 Series Oscilloscopes User’s GuideSaving and Recalling Waveforms and SetupsUsing the oscillosco

Pagina 6

Saving, Recalling, and Printing Data 5Agilent 3000 Series Oscilloscopes User’s Guide 103To save and load oscilloscope setupsYou can save/load oscillo

Pagina 7

5 Saving, Recalling, and Printing Data104 Agilent 3000 Series Oscilloscopes User’s GuideSaving to and Recalling from a USB Mass Storage DeviceIf the

Pagina 8

Saving, Recalling, and Printing Data 5Agilent 3000 Series Oscilloscopes User’s Guide 105For information on using the Save menu, see “To save a file t

Pagina 9

5 Saving, Recalling, and Printing Data106 Agilent 3000 Series Oscilloscopes User’s GuideTo save data to CSV format files (on a USB device)You can sav

Pagina 10 - Contents

Saving, Recalling, and Printing Data 5Agilent 3000 Series Oscilloscopes User’s Guide 107In the Save menu:• To select the folder in which you want to

Pagina 11

5 Saving, Recalling, and Printing Data108 Agilent 3000 Series Oscilloscopes User’s GuideTo load a file from a USB deviceWhen loading a waveform or se

Pagina 12

Saving, Recalling, and Printing Data 5Agilent 3000 Series Oscilloscopes User’s Guide 109Printing the Display to a USB PrinterYou need a USB cable to

Pagina 13

ContentsAgilent 3000 Series Oscilloscopes User’s Guide 11 Returning the Oscilloscope to Agilent for Service 146Contacting Agilent 146ASafety Notices14

Pagina 14

5 Saving, Recalling, and Printing Data110 Agilent 3000 Series Oscilloscopes User’s GuideSupported PrintersHP DeskJet and LaserJet printers are suppor

Pagina 15

A111Agilent 3000 Series OscilloscopesUser’s and Service Guide 6Oscilloscope Utility SettingsPerforming Mask Tests 112Setting IO Parameters 115Settin

Pagina 16

6 Oscilloscope Utility Settings112 Agilent 3000 Series Oscilloscopes User’s GuidePerforming Mask TestsThe mask test function monitors waveform change

Pagina 17 - Getting Started

Oscilloscope Utility Settings 6Agilent 3000 Series Oscilloscopes User’s Guide 113To turn on/off the mask test message display1 In the Mask Test menu

Pagina 18

6 Oscilloscope Utility Settings114 Agilent 3000 Series Oscilloscopes User’s GuideTo set up masksYou can create masks by adding horizontal and vertica

Pagina 19 - Figure 1 Package Contents

Oscilloscope Utility Settings 6Agilent 3000 Series Oscilloscopes User’s Guide 115Setting IO ParametersWhen programming the oscilloscope over the RS-

Pagina 20 - Table 1 Power Cords

6 Oscilloscope Utility Settings116 Agilent 3000 Series Oscilloscopes User’s GuideTurning Sound ON or OFFTo turn the oscilloscope’s beeper sound on or

Pagina 21

Oscilloscope Utility Settings 6Agilent 3000 Series Oscilloscopes User’s Guide 117Performing Self-TestsThe oscilloscope’s Self- Test menu lets you per

Pagina 22

6 Oscilloscope Utility Settings118 Agilent 3000 Series Oscilloscopes User’s GuideRunning Self-CalibrationThe automatic calibration routine adjusts th

Pagina 23 - Step 4. Input a waveform

A119Agilent 3000 Series OscilloscopesUser’s and Service Guide 7Specifications and CharacteristicsSpecifications 120Characteristics 121This chapter d

Pagina 24

12 Agilent 3000 Series Oscilloscopes User’s Guide Contents

Pagina 25

7 Specifications and Characteristics120 Agilent 3000 Series Oscilloscopes User’s GuideSpecificationsNOTEAll specifications are warranted. Specificati

Pagina 26 - Using the Oscilloscope Menus

Specifications and Characteristics 7Agilent 3000 Series Oscilloscopes User’s Guide 121CharacteristicsNOTEAll characteristics are the typical performa

Pagina 27

7 Specifications and Characteristics122 Agilent 3000 Series Oscilloscopes User’s GuideDynamic range: ±6 divInput resistance: 1 MΩ ± 1%Input capacitan

Pagina 28 - Figure 5 Oscilloscope Display

Specifications and Characteristics 7Agilent 3000 Series Oscilloscopes User’s Guide 123Selections:• Edge Trigger on a rising or falling edge of any so

Pagina 29 - Step 7. Use Auto-Scale

7 Specifications and Characteristics124 Agilent 3000 Series Oscilloscopes User’s GuideTable 9 Display System CharacteristicsName Typical ValueDisplay

Pagina 30

Specifications and Characteristics 7Agilent 3000 Series Oscilloscopes User’s Guide 125Table 12 Power RequirementsName Typical ValueLine voltage: Rang

Pagina 31 - Step 8. Compensate probes

7 Specifications and Characteristics126 Agilent 3000 Series Oscilloscopes User’s Guide

Pagina 32 - High Frequency Compensation

A127Agilent 3000 Series OscilloscopesUser’s and Service Guide 8ServiceTesting Performance 128Vertical Performance Verification 130Performance Test R

Pagina 33

8Service128 Agilent 3000 Series Oscilloscopes User’s GuideTesting PerformanceThis section documents performance test procedures. Performance verifica

Pagina 34

Service 8Agilent 3000 Series Oscilloscopes User’s Guide 129Before Testing PerformanceEquipment RequiredCalibration1 Push the Utility button on the fr

Pagina 35 - Displaying Data

Agilent 3000 Series Oscilloscopes User’s Guide 13 FiguresFigure 1. Package Contents 19Figure 2. Save/Recall Button 22Figure 3. Front Panel 24Figure 4.

Pagina 36 - Using the Horizontal Controls

8Service130 Agilent 3000 Series Oscilloscopes User’s GuideVertical Performance VerificationThis section contains the following vertical performance v

Pagina 37 - Displaying Data 2

Service 8Agilent 3000 Series Oscilloscopes User’s Guide 131Procedure1 Disconnect all cables from the oscilloscope channel inputs.2 Press Save/Recall

Pagina 38 - Slow Scan Mode

8Service132 Agilent 3000 Series Oscilloscopes User’s Guide7 Select the Averages item in the Acquire menu until “256” appears.8 Set the channel 1 vert

Pagina 39

Service 8Agilent 3000 Series Oscilloscopes User’s Guide 13311 Press Measure.12 Select the Voltage menu item.13 Select the Vavg measurement as shown

Pagina 40 - Zoomed window view

8Service134 Agilent 3000 Series Oscilloscopes User’s Guide14 For each channel 1 vertical sensitivity in the DC Gain Test section of the “Performance

Pagina 41 - X-Y Format

Service 8Agilent 3000 Series Oscilloscopes User’s Guide 13519 Select Default Setup in the Save/Recall menu.20 Set the channel 2 vertical sensitivity

Pagina 42 - To specify a trigger holdoff

8Service136 Agilent 3000 Series Oscilloscopes User’s GuideSpecificationEquipment RequiredConnectionsConnect the equipment as shown in Figure 56.Table

Pagina 43 - To view the sample rate

Service 8Agilent 3000 Series Oscilloscopes User’s Guide 137Procedure1 Preset and calibrate the power meter according to the instructions found in the

Pagina 44 - Using the Vertical Controls

8Service138 Agilent 3000 Series Oscilloscopes User’s Guide5 Select the Default Setup item in the Save/Recall menu.6 Press the Auto-Scale front panel

Pagina 45 - To adjust the vertical scale

Service 8Agilent 3000 Series Oscilloscopes User’s Guide 1398 Set the horizontal scale to 500 ns/div.Figure 58 Channel 1 Vertical Scale SettingFigure

Pagina 46 - To specify channel coupling

14 Agilent 3000 Series Oscilloscopes User’s Guide FiguresFigure 29. Sample Rate and Oscilloscope Bandwidth 65Figure 30. Acquire Button 68Figure 31. Re

Pagina 47 - AC Coupling Status

8Service140 Agilent 3000 Series Oscilloscopes User’s Guide9 Press Acquire.10 Select the Mode menu item until “Average” appears.11 Select the Average

Pagina 48 - To specify a bandwidth limit

Service 8Agilent 3000 Series Oscilloscopes User’s Guide 14116 Using the Vpp reading, calculate the Vrms value using the following expression and reco

Pagina 49 - Bandwidth limit OFF

8Service142 Agilent 3000 Series Oscilloscopes User’s Guide18 Calculate the reference gain as follows:Record this value in the Calculated Gain @ 1 MHz

Pagina 50 - Probe Attenuation

Service 8Agilent 3000 Series Oscilloscopes User’s Guide 14323 Calculate the gain at the maximum frequency using the expression and record it in the “

Pagina 51 - To invert a waveform

8Service144 Agilent 3000 Series Oscilloscopes User’s GuidePerformance Test RecordTable 21 DC Gain TestVertical SensitivityPower Supply SettingVDMM+ V

Pagina 52 - 2 Displaying Data

Service 8Agilent 3000 Series Oscilloscopes User’s Guide 145Channel 22mV/div ±6mV +0.96 to +1.045mV/div ±15mV +0.96 to +1.0410 mV/div ±30 mV +0.97 to

Pagina 53 - Using Math Function Waveforms

8Service146 Agilent 3000 Series Oscilloscopes User’s GuideReturning the Oscilloscope to Agilent for ServiceBefore shipping the oscilloscope to Agilen

Pagina 54

A147Agilent 3000 Series OscilloscopesUser’s and Service Guide ASafety NoticesWarnings 147To clean the instrument 148Safety Symbols 149This

Pagina 55

ASafety Notices 148 Agilent 3000 Series Oscilloscopes User’s Guide• Whenever it is likely that the ground protection is impaired, you must make the

Pagina 56 - FFT Resolution

Safety Notices A Agilent 3000 Series Oscilloscopes User’s Guide 149Safety SymbolsInstruction manual symbol: the product is marked with this symbol wh

Pagina 57 - To save a reference waveform

Agilent 3000 Series Oscilloscopes User’s Guide 15 TablesTable 1. Power Cords 20Table 2. Auto-Scale Default Settings 30Table 3. FFT Window Characterist

Pagina 58 - To change the grid

ASafety Notices 150 Agilent 3000 Series Oscilloscopes User’s Guide

Pagina 59 - To clear the display

Agilent 3000 Series Oscilloscopes User’s Guide 151 IndexNumerics50% trigger level, 76AAC channel coupling, 46, 54AC trigger coupling, 80, 83AC trigger

Pagina 60 - To invert screen colors

152 Agilent 3000 Series Oscilloscopes User’s GuideIndex external trigger input, 83Ffactory default, 22Fall Time measurement, 94FFT (Fast Fourier Trans

Pagina 61 - Capturing Data

Agilent 3000 Series Oscilloscopes User’s Guide 153Index non-repetitive waveforms, 69nonvolatile memory locations, 101Normal acquisition mode, 72normal

Pagina 62 - Aliasing

154 Agilent 3000 Series Oscilloscopes User’s GuideIndex Ttemperature at last calibration, 120test equipment, 128testing performance, 128theory, sampli

Pagina 64

A Agilent Technologies, Inc. 2005, 2007-2008Available in electronic format onlyFourth Edition, September 2008*D3000-97017*D3000-97017www.agilent.com

Pagina 65 - Oscilloscope Rise Time

16 Agilent 3000 Series Oscilloscopes User’s Guide Tables

Pagina 66

A17Agilent 3000 Series OscilloscopesUser’s and Service Guide 1Getting StartedStep 1. Inspect the package contents 18Step 2. Turn on the oscilloscope

Pagina 67 - Memory Depth and Sample Rate

1Getting Started18 Agilent 3000 Series Oscilloscopes User’s GuideStep 1. Inspect the package contents1 Inspect the shipping container for damage.Keep

Pagina 68 - Choosing the Sampling Mode

Getting Started 1Agilent 3000 Series Oscilloscopes User’s Guide 19sCD-ROM ManualsFigure 1 Package Contents

Pagina 69

Agilent 3000 Series Oscilloscopes User’s GuideNotices© Agilent Technologies, Inc. 2005, 2007-2008No part of this manual may be reproduced in any form

Pagina 70

1Getting Started20 Agilent 3000 Series Oscilloscopes User’s GuideTable 1 Power CordsOption Country Cable Part No.Option 900 United Kingdom, Cyprus, N

Pagina 71

Getting Started 1Agilent 3000 Series Oscilloscopes User’s Guide 21Step 2. Turn on the oscilloscopeThe next few steps (turning on the oscilloscope, lo

Pagina 72 - Choosing the Acquisition Mode

1Getting Started22 Agilent 3000 Series Oscilloscopes User’s GuideStep 3. Load the default oscilloscope setupYou can recall the factory default setup

Pagina 73 - Capturing Data 3

Getting Started 1Agilent 3000 Series Oscilloscopes User’s Guide 23Step 4. Input a waveform1 Input a waveform to a channel of the oscilloscope.Use one

Pagina 74

1Getting Started24 Agilent 3000 Series Oscilloscopes User’s GuideStep 5. Become familiar with the Front Panel ControlsBefore using the oscilloscope,

Pagina 75

Getting Started 1Agilent 3000 Series Oscilloscopes User’s Guide 25The definitions of the buttons and the knobs are as follows:Measure controls Cursor

Pagina 76 - To adjust the trigger level

1Getting Started26 Agilent 3000 Series Oscilloscopes User’s GuideUsing the Oscilloscope MenusWhen one of the oscilloscope front panel buttons turns o

Pagina 77 - To force a trigger

Getting Started 1Agilent 3000 Series Oscilloscopes User’s Guide 27The Menu On/Off button turns off the menu or turns on the last accessed menu again

Pagina 78 - To set up edge triggers

1Getting Started28 Agilent 3000 Series Oscilloscopes User’s GuideStep 6. Become familiar with the oscilloscope displayChannel 1 StatusTimebase Status

Pagina 79

Getting Started 1Agilent 3000 Series Oscilloscopes User’s Guide 29Step 7. Use Auto-ScaleThe oscilloscope has an auto- scale feature that automaticall

Pagina 80 - To set up video triggers

Agilent 3000 Series Oscilloscopes User’s Guide 3 Agilent 3000 Series Oscilloscopes—At a GlanceThe Agilent 3000 Series oscilloscopes are low-cost porta

Pagina 81

1Getting Started30 Agilent 3000 Series Oscilloscopes User’s GuideThe oscilloscope is configured to the following default control settings:Table 2 Aut

Pagina 82 - To set the trigger coupling

Getting Started 1Agilent 3000 Series Oscilloscopes User’s Guide 31Step 8. Compensate probesCompensate probes to match your probe to the input channel

Pagina 83 - To record waveforms

1Getting Started32 Agilent 3000 Series Oscilloscopes User’s GuideHigh Frequency CompensationFor the supplied passive probes:1 Using the BNC adapter,

Pagina 84 - To play-back waveforms

Getting Started 1Agilent 3000 Series Oscilloscopes User’s Guide 33Step 9. Use the Run Control buttonsThere are two buttons for starting and stopping

Pagina 85

1Getting Started34 Agilent 3000 Series Oscilloscopes User’s Guide

Pagina 86 - To store recorded waveforms

A35Agilent 3000 Series OscilloscopesUser’s and Service Guide 2Displaying DataUsing the Horizontal Controls 36Using the Vertical Controls 44Using Mat

Pagina 87 - Making Measurements

2 Displaying Data36 Agilent 3000 Series Oscilloscopes User’s GuideUsing the Horizontal ControlsThe horizontal controls consist of:• The horizontal sc

Pagina 88

Displaying Data 2Agilent 3000 Series Oscilloscopes User’s Guide 37Horizontal scalesettingSamplerateDisplayed waveform window ([ ])position in memoryT

Pagina 89

2 Displaying Data38 Agilent 3000 Series Oscilloscopes User’s GuideTo adjust the horizontal scale• Turn the horizontal scale knob to change the horizo

Pagina 90 - Vpp (Peak-to-Peak Voltage)

Displaying Data 2Agilent 3000 Series Oscilloscopes User’s Guide 39To adjust the horizontal position• Turn the horizontal position knob to change the

Pagina 91

4 Agilent 3000 Series Oscilloscopes User’s Guide In This BookThis guide shows how to use the Agilent 3000 Series oscilloscopes.1Getting StartedDescrib

Pagina 92 - Preshoot

2 Displaying Data40 Agilent 3000 Series Oscilloscopes User’s GuideTo change the horizontal time base (Y-T or X-Y)1 Press Main/Delayed.2 In the Main/

Pagina 93 - Frequency

Displaying Data 2Agilent 3000 Series Oscilloscopes User’s Guide 41X-Y FormatThis format compares the voltage level of two waveforms point by point. I

Pagina 94 - Fall Time

2 Displaying Data42 Agilent 3000 Series Oscilloscopes User’s Guide• Horizontal position knob.• Trigger controls.To reset the trigger offsetTurning th

Pagina 95 - Negative Duty Cycle

Displaying Data 2Agilent 3000 Series Oscilloscopes User’s Guide 43To specify a trigger holdoff:1 Press Main/Delayed.2 In the Main/Delayed menu, sele

Pagina 96 - Counter (Frequency)

2 Displaying Data44 Agilent 3000 Series Oscilloscopes User’s GuideUsing the Vertical ControlsThe vertical controls consist of:• The channel (1, 2),

Pagina 97 - Making Measurements 4

Displaying Data 2Agilent 3000 Series Oscilloscopes User’s Guide 45To turn waveforms on or off (channel, math, or reference)In these situations, press

Pagina 98 - Making Cursor Measurements

2 Displaying Data46 Agilent 3000 Series Oscilloscopes User’s GuideTo adjust the vertical positionAdjusting their vertical position lets you compare w

Pagina 99

Displaying Data 2Agilent 3000 Series Oscilloscopes User’s Guide 47Figure 16 DC Coupling ControlFigure 17 AC Coupling ControlDC CouplingDC Coupling St

Pagina 100

2 Displaying Data48 Agilent 3000 Series Oscilloscopes User’s GuideTo specify a bandwidth limitWhen high frequency components of a waveform are not im

Pagina 101

Displaying Data 2Agilent 3000 Series Oscilloscopes User’s Guide 49Figure 19 BW Limit Control ONFigure 20 BW Limit Control OFF20 MHz BandwidthBandwidt

Pagina 102 - To save and load waveforms

Agilent 3000 Series Oscilloscopes User’s Guide 5 ContentsAgilent 3000 Series Oscilloscopes—At a Glance 3In This Book 4Figures 13Tables 151 Getting Sta

Pagina 103

2 Displaying Data50 Agilent 3000 Series Oscilloscopes User’s GuideTo specify the probe attenuationFor correct measurements, you must match the oscill

Pagina 104

Displaying Data 2Agilent 3000 Series Oscilloscopes User’s Guide 51To use a digital filterYou can apply a digital filter to the sampled waveform data.

Pagina 105

2 Displaying Data52 Agilent 3000 Series Oscilloscopes User’s GuideFigure 22 Waveform Before InversionFigure 23 Waveform After Inversion

Pagina 106

Displaying Data 2Agilent 3000 Series Oscilloscopes User’s Guide 53Using Math Function WaveformsThe math functions control allows the selection of the

Pagina 107

2 Displaying Data54 Agilent 3000 Series Oscilloscopes User’s GuideTo add, subtract, or multiply waveforms1 Press Math.2 In the Math menu, press Oper

Pagina 108

Displaying Data 2Agilent 3000 Series Oscilloscopes User’s Guide 554 Press Window until the desired window is selected:There are four FFT windows. Ea

Pagina 109

2 Displaying Data56 Agilent 3000 Series Oscilloscopes User’s GuideFigure 25 FFT WaveformNOTEFFT ResolutionThe FFT resolution is the quotient of the s

Pagina 110 - Supported Printers

Displaying Data 2Agilent 3000 Series Oscilloscopes User’s Guide 57Using a Reference WaveformYou can save a reference waveform to an internal, nonvola

Pagina 111 - Oscilloscope Utility Settings

2 Displaying Data58 Agilent 3000 Series Oscilloscopes User’s GuideChanging the Display SettingsTo display waveforms as vectors or dots1 Press Displa

Pagina 112 - To run/stop a mask test

Displaying Data 2Agilent 3000 Series Oscilloscopes User’s Guide 59To use waveform persistence1 Press Display.2 In the Display menu, select Persist t

Pagina 113

6 Agilent 3000 Series Oscilloscopes User’s Guide Contents2 Displaying Data 35Using the Horizontal Controls 36To adjust the horizontal scale 38To adjus

Pagina 114 - To set up masks

2 Displaying Data60 Agilent 3000 Series Oscilloscopes User’s GuideTo invert screen colors1 Press Display.2 In the Display menu, select Screen to tog

Pagina 115 - Setting the Menu Language

A61Agilent 3000 Series OscilloscopesUser’s and Service Guide 3Capturing DataOverview of Sampling 62Choosing the Sampling Mode 68Choosing the Acquisi

Pagina 116 - Displaying System Information

3Capturing Data62 Agilent 3000 Series Oscilloscopes User’s GuideOverview of SamplingTo understand the oscilloscope’s sampling and acquisition modes,

Pagina 117 - To perform key tests

Capturing Data 3Agilent 3000 Series Oscilloscopes User’s Guide 63Oscilloscope Bandwidth and Sample RateAn oscilloscope’s bandwidth is typically descr

Pagina 118 - Running Self-Calibration

3Capturing Data64 Agilent 3000 Series Oscilloscopes User’s GuideHowever, digital signals have frequency components above the fundamental frequency (s

Pagina 119

Capturing Data 3Agilent 3000 Series Oscilloscopes User’s Guide 65So, in practice, an oscilloscope’s sample rate should be four or more times its band

Pagina 120 - Specifications

3Capturing Data66 Agilent 3000 Series Oscilloscopes User’s GuideAn oscilloscope’s rise time is not the fastest edge speed that the oscilloscope can a

Pagina 121 - Characteristics

Capturing Data 3Agilent 3000 Series Oscilloscopes User’s Guide 67Memory Depth and Sample RateThe number of points of oscilloscope memory is fixed, an

Pagina 122

3Capturing Data68 Agilent 3000 Series Oscilloscopes User’s GuideChoosing the Sampling ModeThe osilloscope can operate in real- time or equivalent- ti

Pagina 123

Capturing Data 3Agilent 3000 Series Oscilloscopes User’s Guide 69To select the real-time sampling modeIn the real- time sampling mode, single wavefor

Pagina 124 - →2 ), Hardware Counter

ContentsAgilent 3000 Series Oscilloscopes User’s Guide 7 Changing the Display Settings 58To display waveforms as vectors or dots 58To change the grid

Pagina 125 - Table 14 Other

3Capturing Data70 Agilent 3000 Series Oscilloscopes User’s GuideTo select the equivalent-time sampling modeIn the equivalent- time sampling mode (als

Pagina 126

Capturing Data 3Agilent 3000 Series Oscilloscopes User’s Guide 71To select the Equivalent- Time sampling mode:1 Press Acquire.2 In the Acquire menu,

Pagina 127

3Capturing Data72 Agilent 3000 Series Oscilloscopes User’s GuideChoosing the Acquisition ModeThe osilloscope can operate in normal, average, or peak

Pagina 128 - Testing Performance

Capturing Data 3Agilent 3000 Series Oscilloscopes User’s Guide 73The Average acquisition mode decreases the screen refresh rate.Figure 33 Noisy Wavef

Pagina 129 - Before Testing Performance

3Capturing Data74 Agilent 3000 Series Oscilloscopes User’s GuideTo select the Average acquisition mode:1 Press Acquire.2 In the Acquire menu, press

Pagina 130 - DC Gain Accuracy Test

Capturing Data 3Agilent 3000 Series Oscilloscopes User’s Guide 75Because minimum and maximum values for a sample period are stored, you can use the P

Pagina 131 - Procedure

3Capturing Data76 Agilent 3000 Series Oscilloscopes User’s GuideAdjusting the Trigger LevelTo adjust the trigger level• Turn the trigger Level knob.T

Pagina 132

Capturing Data 3Agilent 3000 Series Oscilloscopes User’s Guide 77To force a triggerTo make an acquisition even if no valid trigger has been found:1 P

Pagina 133

3Capturing Data78 Agilent 3000 Series Oscilloscopes User’s GuideChoosing the Trigger ModeThe trigger determines when captured data should be stored a

Pagina 134

Capturing Data 3Agilent 3000 Series Oscilloscopes User’s Guide 794 Press Slope to select the edge to trigger on:• — Rising edge.• — Falling edge.T

Pagina 135

8 Agilent 3000 Series Oscilloscopes User’s Guide ContentsSetting Other Trigger Parameters 82To set the trigger sweep 82To set the trigger coupling 82U

Pagina 136 - Connections

3Capturing Data80 Agilent 3000 Series Oscilloscopes User’s GuideTo set up video triggersVideo triggering is used to trigger on fields or lines of NTS

Pagina 137

Capturing Data 3Agilent 3000 Series Oscilloscopes User’s Guide 81Figure 37 Line SynchronizationFigure 38 Field Synchronization

Pagina 138

3Capturing Data82 Agilent 3000 Series Oscilloscopes User’s GuideSetting Other Trigger ParametersThese are trigger system parameters that apply in all

Pagina 139 - Service 8

Capturing Data 3Agilent 3000 Series Oscilloscopes User’s Guide 832 In the Trigger menu, when “Edge” or “Pulse” is selected for Mode, press Coupling t

Pagina 140

3Capturing Data84 Agilent 3000 Series Oscilloscopes User’s GuideTo select the source channel for recording1 In the Sequence menu (Acquire→Sequence→Mo

Pagina 141 - Vpp reading

Capturing Data 3Agilent 3000 Series Oscilloscopes User’s Guide 85To play-back/stop the recording1 In the Sequence menu (Acquire→Sequence→Mode=Play ba

Pagina 142

3Capturing Data86 Agilent 3000 Series Oscilloscopes User’s GuideTo store recorded waveformsTo store recorded waveforms:1 Press Acquire.2 In the Acqu

Pagina 143

A87Agilent 3000 Series OscilloscopesUser’s and Service Guide 4Making MeasurementsDisplaying Automatic Measurements 88Voltage Measurements 90Time Mea

Pagina 144 - Performance Test Record

4 Making Measurements88 Agilent 3000 Series Oscilloscopes User’s GuideDisplaying Automatic MeasurementsYou can use the Measure button to display auto

Pagina 145

Making Measurements 4Agilent 3000 Series Oscilloscopes User’s Guide 89To clear automatic measurements from the display1 Press Measure.2 In the Measu

Pagina 146 - Contacting Agilent

ContentsAgilent 3000 Series Oscilloscopes User’s Guide 9 Time Measurements 93Frequency 93Period 93Rise Time 94Fall Time 94Positive Pulse Width 95Negat

Pagina 147 - Warnings

4 Making Measurements90 Agilent 3000 Series Oscilloscopes User’s GuideVoltage MeasurementsThere are 10 automatic voltage measurements:• Vpp (Peak- to

Pagina 148 - To clean the instrument

Making Measurements 4Agilent 3000 Series Oscilloscopes User’s Guide 91Vmax (Maximum Voltage)The maximum amplitude. The most positive peak voltage mea

Pagina 149 - Safety Symbols

4 Making Measurements92 Agilent 3000 Series Oscilloscopes User’s GuideVrms (Root-Mean-Square Voltage)The true root- mean- square voltage over the ent

Pagina 150 - ASafety Notices

Making Measurements 4Agilent 3000 Series Oscilloscopes User’s Guide 93Time MeasurementsThere are 10 automatic time measurements plus the hardware fre

Pagina 151

4 Making Measurements94 Agilent 3000 Series Oscilloscopes User’s GuideRise TimeMeasures the rise time of a waveform.Fall TimeMeasures the fall time o

Pagina 152

Making Measurements 4Agilent 3000 Series Oscilloscopes User’s Guide 95Positive Pulse WidthMeasures the positive pulse width of a waveform.Negative Pu

Pagina 153

4 Making Measurements96 Agilent 3000 Series Oscilloscopes User’s GuideDelay Between Rising EdgesMeasures the delay between two waveforms using the ri

Pagina 154

Making Measurements 4Agilent 3000 Series Oscilloscopes User’s Guide 97The counter uses the trigger comparator to count the number of cycles within a

Pagina 155

4 Making Measurements98 Agilent 3000 Series Oscilloscopes User’s GuideMaking Cursor MeasurementsYou can use the Cursors button to select between thes

Pagina 156 - *D3000-97017*

Making Measurements 4Agilent 3000 Series Oscilloscopes User’s Guide 995 To adjust the cursors:• Select CurA and turn the Entry knob to adjust the “A