Agilent Technologies B1500A Manuale di Servizio Pagina 414

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 598
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 413
4-98 Agilent B1500A/B1505A Programming Guide, Edition 11
Command Reference
ERPFUHCA?
Example
Statements
OUTPUT @B1500;"ERPFUHCA 3,4"
ERPFUHCA?
This query command returns the slot numbers for the MC/HCSMU modules
connected to the V Control and I Control inputs of Agilent N1265A ultra high
current expander/fixture.
Syntax ERPFUHCA?
Query Response vsmu,ismu<CR/LF^EOI>
vsmu : Slot number for the MC/HCSMU connected to the V Control input.
ismu : Slot number for the MC/HCSMU connected to the I Control input.
Example
Statements
OUTPUT @B1500;"ERPFUHCA?"
ENTER @B1500;vsmuUHC,ismuUHC
ERPFUHCCAL?
This query command performs the self-calibration of Agilent N1265A ultra high
current expander/fixture, and returns the execution results. After this command, read
the result soon.
Execution
Conditions
The ERPFUHCA command has been executed before this command.
Syntax ERPFUHCCAL?
Query Response result<CR/LF^EOI>
0: Pass
1: Fail
Example
Statements
OUTPUT @B1500;"ERPFUHCCAL?"
ENTER @B1500;result
ERPFUHCMAX?
This query command returns the available maximum current for the
currently-connected Agilent N1265A ultra high current expander/fixture.
Vedere la pagina 413
1 2 ... 409 410 411 412 413 414 415 416 417 418 419 ... 597 598

Commenti su questo manuale

Nessun commento