Agilent 75000 Series B
Agilent E1351A/52A/53A/57A/58A
FET Multiplexers
Service Manual
Enclosed is the Service Manual for the Agilent E1351A/52A/53A/57A/58A
FET Multiplexers. Insert this manual, along with any other VXIbus manuals
that you have, into the binder that came with your Agilent Mainframe.
*E1351-90011*
Manual Part Number: E1351-90011
Printed in Malaysia E0912
Vedere la pagina 0
1 2 3 4 5 6 ... 67 68

Sommario

Pagina 1 - *E1351-90011*

Agilent 75000 Series BAgilent E1351A/52A/53A/57A/58AFET MultiplexersService Manual Enclosed is the Service Manual for the Agilent E1351A/

Pagina 2

Chapter 3 — Replaceable partsIntroduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Pagina 3

Chapter 1General InformationIntroduction This manual contains information required to test, troubleshoot, and repairthe Agilent E1351A, E1352A, E1353A

Pagina 4 - Safety Symbols

SafetyConsiderationsThis product is a Safety Class I instrument that is provided with a protectiveearth terminal when installed in the mainframe. Chec

Pagina 5 - Declaration of Conformity

WARNINGUSING AUTOTRANSFORMERS. If the mainframe is to beenergized via an autotransformer (for voltage reduction) makesure the common terminal is conne

Pagina 6

FET Multiplexer DescriptionThe Agilent E1351A, E1352A, E1353A, E1357A, and E1358A FETMultiplexer is an "instrument" in a VXIbus mainframe. A

Pagina 7 - Manual Descriptions

Agilent E1357ADescriptionThe Agilent E1357A provides up to eight channels of strain gage switching.Strain gage measurements are supported in 1/4 bridg

Pagina 8 - What’s in this Manual

RecommendedTest EquipmentTable 1-1 lists the test equipment recommended for testing, adjusting, andservicing the FET Multiplexers. Essential requireme

Pagina 9 - Table of Contents

Notify AgilentNotify Agilent and carrier.Figure 1-2. Initial (Incoming) Inspection GuidelinesGeneral Information 1-7

Pagina 10 - Chapter 4 — Service

ShippingGuidelinesFollow the steps in Figure 1-3 to return a FET Multiplexer to an AgilentTechnologies Sales and Support Office or Service Center. 1

Pagina 11 - General Information

Chapter 2Verification TestsIntroduction This chapter describes the verification tests for the Agilent E1351A,E1352A, E1353A, E1357A, and E1358A. The t

Pagina 13 - General Information 1-3

FunctionalVerification TestThe Functional Verification Test for the Agilent E1351A, E1352A,E1353A, E1357A, and E1358A FET Multiplexers consists of sen

Pagina 14 - Description

OperationVerification TestThe procedures in this section are used to provide a high level of confidencethat the FET Multiplexer is meeting published s

Pagina 15 - General Information 1-5

re-wire each time the tests are performed. The Agilent E1351A terminalblock and case assembly part number is E1351-80001.Test 2-1: ClosedChannelResist

Pagina 16 - Shipping

• Repeat step 3 for channels 01 through 15 • Use CLOS (@nncc), where nn is the card number and cc is thechannel numberLO ChannelMeasurements 1. Make

Pagina 17 - General Information 1-7

Tree Switch HI ChannelMeasurements 1. Make Hardware Connections• Turn mainframe power OFF• Connect DMM as shown in Figure 2-4• Turn mainframe power O

Pagina 18 - Guidelines

Tree Switch LOChannel Measurements 1. Make Hardware Connections• Turn mainframe power OFF• Connect DMM as shown in Figure 2-5• Turn mainframe power O

Pagina 19 - Verification Tests

Example: ClosedChannel ResistanceTestThis example performs a closed channel resistance test of all measurementpaths. If a FET on resistance is >3.1

Pagina 20 - Verification

370 OUTPUT @Dmm;"TRIG SGL"380 ENTER @Dmm;Value390 IF Value<10000 THEN400 CLEAR SCREEN410 PRINT "Measurement indi

Pagina 21 - Verification Tests 2-3

780 PAUSE790 OUTPUT @Dmm;"PRESET NORM;FUNC OHM"800 OUTPUT @Mux;"*RST"810 !820 ! Check for stuck tree switches8

Pagina 22 - Resistance Test

Typical ResultClosed channel resistance measurement resultsChannel 0 HI 989 Ohms LO 1004 OhmsChannel 1 HI 991 Ohms LO 979 OhmsChannel 2 HI 1001 Ohms L

Pagina 23 - Verification Tests 2-5

CertificationAgilent Technologies certifies that this product met its published specifications at the time of shipment from the factory. AgilentTechn

Pagina 24 - 2-6 Verification Tests

Test 2-2: LeakageTestThe test verifies the input impedance by measuring the voltage drop across aknown resistor in series with the input impedance. Le

Pagina 25 - Verification Tests 2-7

3. Check Channels Leakage • Send CLOS(@nn00) to the FET Multiplexer to close channel00, where nn is card number (typically 01)• Send TRIG SGL to DMM•

Pagina 26 - 2-8 Verification Tests

HI to Chassis Leakage 1. Make Hardware Connections• Turn power supply and mainframe power OFF• Connect DMM, power supply, and resistor as shown in F

Pagina 27 - Verification Tests 2-9

3. Check Channels Leakage • Send CLOS(@nn00) to the FET Multiplexer, where nn is thechannel number (typically 01)• Send TRIG SGL to DMM• Record the D

Pagina 28 - 2-10 Verification Tests

• Send CLOS(@nn15) to the FET Multiplexer to close channel15 and tree B• Send TRIG SGL to DMM• Record the DMM reading in Table 2-1 Positive Polarity,

Pagina 29 - Verification Tests 2-11

LO to Chassis Leakage 1. Make Hardware Connections• Turn power supply and mainframe power OFF• Connect DMM, power supply, and resistor as shown in

Pagina 30 - Test 2-2: Leakage

• Record the DMM reading in Table 2-1 Positive Polarity, LO to Chassis, Channels• Send OPEN (@nn00) to the FET Multiplexer4. Check Tree Leakage Curre

Pagina 31 - Verification Tests 2-13

Example: LeakageCurrent TestThis example performs a leakage test from HI to LO, HI to Chassis, and LOto Chassis. If the leakage is too high (caused by

Pagina 32 - 2-14 Verification Tests

170 PRINT " 1. Turn mainframe, power supply, and DMM power OFF" 180 PRINT " 2. Connect GPIB cable between mainframe and DMM

Pagina 33 - Verification Tests 2-15

580 END IF 590 OUTPUT @Mux;"OPEN (@"&Cc$&"00)" 600 OUTPUT @Mux;"CLOS (@"&Cc$&"15)" 61

Pagina 34 - 2-16 Verification Tests

Frame or chassis ground terminal—typicallyconnects to the equipment’s metal frame.Alternating current (AC).Direct current (DC). Indicates hazardous vo

Pagina 35 - Verification Tests 2-17

Typical ResultPositive polarity leakageDirect Channels Tree A Tree BDIRECT HI to DIRECT LO 0.0021 Vdc 0.0015 Vdc 0.0020 Vdc 0.0018 VdcHI to CHASSIS 0.

Pagina 36 - 2-18 Verification Tests

Leakage Test Conditions:• DC Volts function, 100 mV range• 90 day specifications• Worst case reading = 0.01 V• Resistor value: 99 kΩ < R < 101

Pagina 37 - Verification Tests 2-19

Table 2-1. Performance Test Record (Page 1 of 2)General InformationTest Facility:Name _____________________________________Address __________________

Pagina 38 - 2-20 Verification Tests

Table 2-1. Performance Test Record (Page 2 of 3)Test No/Description Minimum*ValueMeasured Value (V) MaximumValue MeasUncert Test AccRatio (TAR)2-1.

Pagina 39 - Verification Tests 2-21

Table 2-1. Performance Test Record (Page 3 of 3)Test No/Description MinimumValue *Measured Value (V) MaximumValue MeasUncert Test AccRatio (TAR)2-2

Pagina 40 - Test Record

Chapter 3Replaceable PartsIntroduction This chapter contains information to order replaceable parts for the AgilentE1351A, E1352A, E1353A, E1357A, and

Pagina 41 - Ratio (TAR)

Table 3-1. FET Multiplexer Replaceable Parts ReferenceDesignatorAgilent PartNumberQty Part Description Mfr.CodeMfr. PartNumberAgilent E1351AA1 E1351-

Pagina 42 - 2-24 Verification Tests

ReferenceDesignatorAgilent PartNumberQty Part Description Mfr.CodeMfr. PartNumberE1351-66201 1 16-CHANNEL FET MULTIPLEXER(See figure 3-1)28480 E1351-6

Pagina 43 - Verification Tests 2-25

Table 3-4. FET Multiplexers Reference Designators FET Multiplexers Reference DesignatorsA ... assembl

Pagina 44 - 2-26 Verification Tests

Mechanical Parts LocatorsFigure 3-1 shows the mechanical replaceable parts on the componentassembly (Agilent PN E1351-66201). Figure 3-2 shows the mec

Pagina 45 - Replaceable Parts

Declaration of Conformity Declarations of Conformity for this product and for other Agilent products may be downloaded from the Internet. There are

Pagina 46 - 3-2 Replaceable Parts

Figure 3-2. Terminal Case Assembly Replaceable Parts3-6 Replaceable Parts

Pagina 47 - Replaceable Parts 3-3

Chapter 4ServiceIntroduction This chapter contains service information for the Agilent E1351A, E1352A,E1353A, E1357A, and E1358A FET Multiplexers. Als

Pagina 48 - 3-4 Replaceable Parts

Figure 4-1. Agilent E1351A Simplified Switch Diagram4-2 Service

Pagina 49 - Locators

Figure 4-2. Agilent E1352A Simplified Switch DiagramService 4-3

Pagina 50 - 3-6 Replaceable Parts

Figure 4-3. Agilent E1353A Simplified Switch Diagram4-4 Service

Pagina 51 - Chapter 4

Figure 4-4. Agilent E1357A/58A Simplified Switch DiagramService 4-5

Pagina 52 - 4-2 Service

for use with the Agilent E1357A and Agilent E1358A Strain Gage FETMultiplexers. Repair Strategy Agilent Technologies recommends replacement of the en

Pagina 53 - Service 4-3

Making VisualChecksVisual checks for the FET Multiplexers include the following. See Table 4-2 for typical checks.• Check switches/jumpers• Check for

Pagina 54 - 4-4 Service

Repair and Maintenance GuidelinesThis section provides guidelines for repairing and maintaining the FETMultiplexer including:• ESD precautions• Solder

Pagina 55 - Service 4-5

Chapter AVerification Tests - C ProgramsFunctionalVerificationTestThis program is designed to do the Functional Verification Test found inChapter 2 -

Pagina 57 - Making Visual

PerformanceVerificationTestsThese programs are designed to do the Performance Verification Testsfound in Chapter 2 - Verification Tests.NOTEThese prog

Pagina 58 - Maintenance

/*...Measure Closed Channel Resistance...*/ iprintf (id, "*RST\n"); iprintf (dm, "PRESET NORM;TRIG

Pagina 59 - Chapter A

printf ("\n\nMeasurements complete for Channel HI and LO"); printf ("\nPress ENTER for Tree Switch measurements"); getchar ();

Pagina 60

if (tree[i][j] > 3100) { if (j == 0) printf ("\n*** Resistance for A Tree Switch is 3.1 kOhms"); else

Pagina 61 - Example C Programs A-3

Example: LeakageCurrent TestThis example performs a leakage test from HI to LO, HI to Chassis, and LOto Chassis. If the leakage is too high (caused b

Pagina 62 - A-4 Example C Programs

if (i == 0) { printf ("\n2. Connect DMM input HI leads to Power Supply HI"); printf ("\n3. Connect DMM input LO leads

Pagina 63 - Example C Programs A-5

if (result[i][1] > .01) printf ("\n\n***** Channel path leakage out of tolerance %6.4lf Volts *****",result[i][1]); /*...

Pagina 64 - Example: Leakage

printf ("\n\n-------------------------------------------------------------"); printf ("\n Negative Polarity Leaka

Pagina 65 - Example C Programs A-7

A-10 Example C Programs

Pagina 66 - A-8 Example C Programs

Agilent 75000 Series B Service DocumentationSuggested Sequence to Use ManualsManual DescriptionsInstallation and Getting Started Guide. This manual

Pagina 67 - Example C Programs A-9

What’s in this ManualManual Overview This manual shows how to service the Agilent E1351A, E1352A, E1353A, E1357A, and E1358A FETMultiplexers. Consult

Pagina 68 - A-10 Example C Programs

Table of ContentsChapter 1 — General InformationIntroduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Commenti su questo manuale

Nessun commento