Agilent Technologies E1356A Manuale di Servizio Pagina 35

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 68
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 34
LO to Chassis Leakage 1. Make Hardware Connections
Turn power supply and mainframe power OFF
Connect DMM, power supply, and resistor as shown in
Figure 2-10
Turn power supply and mainframe power ON
2. Check Direct Terminals Leakage Current
Send *RST to FET Multiplexer
Send TRIG SGL to DMM
Record the DMM reading in Table 2-1
Positive Polarity, LO to Chassis, Direct
The DMM measurement should be less than 0.010 Vdc.
A measurement out of this range indicates a failure of the
FET Multiplexer.
3. Check Channels Leakage
Send CLOS(@nn00) to the FET Multiplexer, where nn is the
card number (typically 01)
Send TRIG SGL to DMM
Agilent E1300B/E1301B
Figure 2-10. Positive LO to Chassis Leakage Connections
Verification Tests 2-17
Vedere la pagina 34
1 2 ... 30 31 32 33 34 35 36 37 38 39 40 ... 67 68

Commenti su questo manuale

Nessun commento