Record the DMM reading in Table 2-1
Positive Polarity, LO to Chassis, Channels
Send OPEN (@nn00) to the FET Multiplexer
4. Check Tree Leakage Current
Send SCAN:PORT ABUS to the FET Multiplexer
Send CLOS(@nn00) to the FET Multiplexer to close channel
00 and Tree A, where nn is the card number (typically 01)
Send TRIG SGL to DMM
Record the DMM reading in Table 2-1
Positive Polarity, LO to Chassis, Tree A
Send OPEN (@nn00) to the FET Multiplexer
Send CLOS(@nn15) to the FET Multiplexer to close channel
15 and Tree B
Send TRIG SGL to DMM
Record the DMM reading in Table 2-1
Positive Polarity, LO to Chassis, Tree B
Send *RST to the FET Multiplexer
5. Change Polarity
Turn power supply and mainframe power OFF
Connect DMM, power supply, and resistor as shown in Figure
2-11
Turn power supply and mainframe power ON
6. Repeat Steps 2 through 4
Record all results in Table 2-1 as
Negative Polarity, LO to Chassis
2-18 Verification Tests