SS1Agilent 81101A 50 MHz Pulse GeneratorReference Guide

Pagina 2

10 About this BookConventions Used in this BookThis book uses certain conventions to indicate elements of the Agilent 81101A’s user interface. The fol

Pagina 3 - Pulse Generator

100 SpecificationsAgilent 81101A SpecificationsTrigger ModesContinuousGenerate continuous pulses, double pulses, or bursts.Externally TriggeredEach ac

Pagina 4 - Warranty

101SpecificationsAgilent 81101A SpecificationsTrigger and Strobe SpecificationsTrigger OutputStrobe OutputTrigger Output Specifications Agilent 81101

Pagina 5 - Services and Support

102 SpecificationsAgilent 81101A SpecificationsTypical DelaysMode from totyp. valueExternal Width Ext Input Strobe/Trigger OutOUTPUT8.5 ns22.5 nsTrigg

Pagina 6 - Safety Summary

103SpecificationsAgilent 81101A SpecificationsHuman InterfaceOverprogrammingParameter values can be entered exceeding the specified range.Warnings an

Pagina 7 - Ground the Instrument

104 SpecificationsAgilent 81101A SpecificationsRemote ControlOperates according to IEEE standard 488.2, 1987 and SCPI 1992.0.Function CodeSH1, AH1, T6

Pagina 8 - Safety Symbols

105SpecificationsPulse Parameter DefinitionsPulse Parameter Definitions Here you find the pulse parameter definitions of terms used in the instrument

Pagina 9 - About this Book

106 SpecificationsPulse Parameter DefinitionsTime Reference PointThe time reference point is at the median of the amplitude (50% amplitude point on pu

Pagina 10 - Conventions Used in this Book

107SpecificationsPulse Parameter Definitionspractice, start points may shift with changes in transition time) when transition times are varied. This

Pagina 11 - Contents

108 SpecificationsPulse Parameter DefinitionsTransition TimeInterval between the 10% and 90% amplitude points on the leading/trailing edge:LinearityPe

Pagina 12

109SpecificationsPulse Parameter DefinitionsJitterShort-term instability of one edge relative to a reference edge. Usually specified as rms value, wh

Pagina 13 - 1General Programming

xiContentsNotice ... 4Safety Summary ...

Pagina 14 - The GP-IB Interface Bus

110 SpecificationsPulse Parameter DefinitionsPreshoot, Overshoot, RingingPreshoot and overshoot are peak distortions preceding/following an edge. Ring

Pagina 15 - Agilent 81101A Remote Control

111SpecificationsPulse Parameter DefinitionsRepeatabilityWhen an instrument operates under the same environmental conditions and with the same settin

Pagina 16 - Programming Recommendations

112 SpecificationsPulse Parameter Definitions

Pagina 17

113IndexSymbols% of Per 57, 58AAcoustic Noise Emission 92Amplitudecurrent 48definition 109voltage 70BBurstlength 84mode 84CCertificationmet specifica

Pagina 18 - Common Command Summary

114 IndexJJitterdefinition 109KKey-Code Reference 80LLast Key Pressed 79Leading Edgecommand 65definition 108Linearitydefinition 108List of Commands 26

Pagina 19 - Status Model

115IndexSSafetysymbols 8SCPICommand Dictionary 38Version 83Settingload into memory ’n’ 46Settling Timedefinition 110Simulating Key Presses 79Skewdefi

Pagina 21 - Enable Register

748 95 61 2 3O . +/-nanomicroMegamillikiloENTERMANSTOREHELPSHIFTDATA ENTRY CURSOR/DIGIT MODIFYLOCALAUTOSETRECALL PERIOD DELAY WIDTHLEAD TRAIL HIGH LOW

Pagina 22 - Standard Event Status Group

Copyright Agilent Technologies 1998, 2000Edition E0300Printed in Germany81101-91021

Pagina 23 - OPERation Status Group

MANUAL CHANGES September, 01Manual for Model Number 81101AManual printed on März

Pagina 24 - QUEStionable Status Group

xii ContentsAgilent 81101A Specifications ... 91General ...

Pagina 25 - 2Programming Reference

MODEL 81101A__________________________________________________________________________________________________________________________________________

Pagina 26 - Agilent 81101A SCPI Command

MODEL 81101A__________________________________________________________________________________________________________________________________________

Pagina 27

MODEL 81101A__________________________________________________________________________________________________________________________________________

Pagina 28

DECLARATION OF CONFORMITY According to ISO/IEC Guide 22 and CEN/CENELEC EN 45014 Revision: C Issue Date: 2001-May-02 Manufacturer’s Name: Ag

Pagina 29

1311General Programming AspectsThis chapter provides general information on writing GP-IB/SCPI programs for instruments like the Agilent 81101A.Detai

Pagina 30

14 General Programming AspectsThe GP-IB Interface BusThe GP-IB Interface BusThe General Purpose Interface Bus is the interface used for communication

Pagina 31 - Programming Reference

15General Programming AspectsAgilent 81101A Remote ControlAgilent 81101A Remote ControlGP-IB Address You can only set the GP-IB address from the fron

Pagina 32

16 General Programming AspectsProgramming RecommendationsProgramming RecommendationsHere are some recommendations for programming the instrument: • St

Pagina 33

17General Programming AspectsProgramming RecommendationsWhen you have found the correct setting, then use this to create the program. In the program

Pagina 34

18 General Programming AspectsCommon Command SummaryCommon Command SummaryThis table summarizes the IEEE 488.2 common commands supported by the Agilen

Pagina 35 - Programming the Instrument

19General Programming AspectsStatus ModelStatus ModelThe instrument has a status reporting system conforming to IEEE 488.2 and SCPI. The above figure

Pagina 36 - Triggered

Per 1.000µµµµs NormalOUTPUT TRG-LEVLIMITSMODE/TRGOFFDelay1 0psWidthLeadETraiE100.0ns 5.00ns =LeadEOffsetAmplit +0.0mV 1.00V50ΩΩΩΩ into 50.0ΩΩΩ

Pagina 37 - External Width

20 General Programming AspectsStatus ModelCondition RegisterA condition register contains the current status of the hardware and firmware. It is conti

Pagina 38 - SCPI Instrument Command List

21General Programming AspectsStatus ModelEnable RegisterThe enable register defines which bits in an event register are included in the logical OR in

Pagina 39

22 General Programming AspectsStatus ModelStatus ByteThe status byte summarizes the information from all other status groups. The summary bit for the

Pagina 40

23General Programming AspectsStatus ModelOPERation Status GroupThis Status Group is not used in the instrument.Bit Description0 Unused, always 01 Unu

Pagina 41

24 General Programming AspectsStatus ModelQUEStionable Status GroupThe QUEStionable Status group is used to report warning conditions amongst the volt

Pagina 42

2522Programming ReferenceThis chapter provides reference information on the following topics: • “Agilent 81101A SCPI Command Summary” on page 26 • “D

Pagina 43 - DISP OFF

26 Programming ReferenceAgilent 81101A SCPI Command SummaryAgilent 81101A SCPI Command SummaryCommand Parameter Descriptionsee page:ARM(Trigger mode a

Pagina 44 - :MMEMory:CDIRectory

27Programming ReferenceAgilent 81101A SCPI Command Summary:OUTPut[1][:NORMal][:STATe]OFF|ON|1|0 Set/read normal output state:IMPedance47[:INTernal]&l

Pagina 45 - :MMEMory:INITialize

28 Programming ReferenceAgilent 81101A SCPI Command Summary:DOUBle[1][:STATe]OFF|ON Enable/disable double pulses per pulse period59:DELay<value>

Pagina 46 - :MMEMory:STORe:STATe

29Programming ReferenceAgilent 81101A SCPI Command Summary:OFFset<value> Set/read channel offset voltage71:HIGH<value> Set/read channel h

Pagina 47 - :OUTPut[1]:IMPedance:EXTernal

Reference GuideAgilent 81101A 50 MHzPulse GeneratorPart No. 81101-91021Printed in Germany March 2000Edition 1.0, E0300

Pagina 48 - :OUTPut[1]:POLarity

30 Programming ReferenceAgilent 81101A SCPI Command Summary[:STATe]ON|OFF Switch security on and off82:SETBlock data Set/read complete instrument sett

Pagina 49

31Programming ReferenceDefault Values, Standard SettingsDefault Values, Standard SettingsParameter *RST, Default Values:ARM : EWIDth:STATe OFF:FREQue

Pagina 50 - High – Low

32 Programming ReferenceDefault Values, Standard Settings:LOW –10.0mA:STATe OFF:FREQ [:CW|:FIXed] 1.00MHz:AUTO not applicable:HOLD VOLT:PHASe[1][:ADJu

Pagina 51

33Programming ReferenceDefault Values, Standard Settings[:SOURce]:VOLTage[1]:LEVel][IMMediate] [:AMPLitude] 1.0V:OFFSet 0.0V:HIGH 500mV:LOW –500mV:LI

Pagina 52

34 Programming ReferenceDefault Values, Standard Settings:LEVel 1.0V:SLOPe POSitive:SOURce IMMediateParameter *RST, Default Values

Pagina 53

35Programming ReferenceProgramming the Instrument Trigger ModesProgramming the Instrument Trigger ModesThe following figure shows the instrument’s ar

Pagina 54 - Frequency

36 Programming ReferenceProgramming the Instrument Trigger ModesUse the :ARM subsystem to select the overall triggering mode of the instrument (CONTIN

Pagina 55 - [:SOURce]:HOLD

37Programming ReferenceProgramming the Instrument Trigger ModesExternal WidthSet External Width mode using the :EWIDth[:STATe] command::ARM:EWIDth ON

Pagina 56 - [:SOURce]:PHASe[1][:ADJust]

38 Programming ReferenceSCPI Instrument Command ListSCPI Instrument Command ListThe following reference sections list the instrument commands in alpha

Pagina 57 - [:SOURce]:PULSe:DELay[1]

39Programming ReferenceSCPI Instrument Command ListCommand :ARM:EWID Long:ARM[:SEQuence[1]|STARt][:LAYer]:EWIDth[:STATe]Form Set & QueryParameter

Pagina 58 - [:SOURce]:PULSe:DELay[1]:HOLD

4 NoticeNoticeCopyright 1998 Agilent Technologies 1998, 2000. All rights reserved. No part of this manual may be reproduced in any form or by any mea

Pagina 59 - [:SOURce]:PULSe:DELay[1]:UNIT

40 Programming ReferenceSCPI Instrument Command ListCommand :ARM:IMPLong:ARM[:SEQuence[1]|STARt][:LAYer]:IMPedanceForm Set & QueryParameter Numeri

Pagina 60 - DblDel% =

41Programming ReferenceSCPI Instrument Command ListCommand :ARM:PERLong:ARM[:SEQuence[1]|STARt][:LAYer]:PERiodForm Set & QueryParameter NumericPa

Pagina 61

42 Programming ReferenceSCPI Instrument Command ListCommand :ARM:SENSLong:ARM[:SEQuence[1]|STARt][:LAYer]:SENSeForm Set & QueryParameterEDGE | LEV

Pagina 62 - Frequency =

43Programming ReferenceSCPI Instrument Command ListCommand :ARM:SOURLong:ARM[:SEQuence[1]|STARt][:LAYer]:SOURceForm Set & QueryParameterIMMediate

Pagina 63 - [:SOURce]:PULSe:TDELay[1]

44 Programming ReferenceSCPI Instrument Command ListCommand :MMEM:CAT?Long:MMEMory:CATalog?Form QueryParameter["A:"]*RST value Not applicabl

Pagina 64

45Programming ReferenceSCPI Instrument Command ListCommand :MMEM:COPYLong:MMEMory:COPYForm EventParameter"filename"[,"A:"],"

Pagina 65

46 Programming ReferenceSCPI Instrument Command ListCommand :MMEM:LOAD:STATLong:MMEMory:LOAD:STATeForm EventParameter<n>,"filename"[,&

Pagina 66

47Programming ReferenceSCPI Instrument Command ListCommand :OUTP[1]Long:OUTPut[1][:NORMal][:STATe]Form Set & QueryParameterON | OFF | 1 | 0*RST v

Pagina 67 - [:AMPlitude]

48 Programming ReferenceSCPI Instrument Command ListSpecified Limits 0.1 Ω to 1 MΩ Description Use this command to set the expected load impedance of

Pagina 68 - [:SOURce]:ROSCillator:SOURce

49Programming ReferenceSCPI Instrument Command ListValue couplingRange coupling OffsetDescription Use this command to program the amplitude current o

Pagina 69

5NoticeAgilent Technologies warrants that its software and firmware designated by Agilent Technologies for use with an instrument will execute its pr

Pagina 70 - Amplitude

50 Programming ReferenceSCPI Instrument Command ListRange coupling AmplitudeDescription Use this command to program the offset current of the OUTPUT s

Pagina 71

51Programming ReferenceSCPI Instrument Command Listhave to execute [:SOURCE]:HOLD CURRent command to enable the [:SOURCE]:CURRent subsystem.The avail

Pagina 72 - Offset =

52 Programming ReferenceSCPI Instrument Command List • Specified Voltage limits • Actual OUTPUT Impedance setting :OUTPut:IMPedance • Actual Expect

Pagina 73 - High–Low

53Programming ReferenceSCPI Instrument Command ListCommand :CURR[1]:LIM:LOWLong[:SOURce]:CURRent[1]:LIMit:LOWForm Set & QueryParameter NumericPar

Pagina 74

54 Programming ReferenceSCPI Instrument Command ListExample To set and activate the current limits for the output:Command :FREQLong[:SOURce]:FREQuency

Pagina 75

55Programming ReferenceSCPI Instrument Command ListCommand :FREQ:AUTOLong[:SOURce]:FREQuency[:CW|:FIXed]:AUTOForm EventParameterONCE*RST value Not ap

Pagina 76 - :STATus:PRESet

56 Programming ReferenceSCPI Instrument Command ListCommand :PHAS[1]Long[:SOURce]:PHASe[1][:ADJust]Form Set & QueryParameter NumericParameter suff

Pagina 77

57Programming ReferenceSCPI Instrument Command ListCommand :PULS:DCYC[1]Long[:SOURce]:PULSe:DCYCle[1]Form Set & QueryParameter NumericValue coupl

Pagina 78 - :SYSTem:ERRor?

58 Programming ReferenceSCPI Instrument Command ListSpecified limits 0.00 ns to 999 s (limited by period – 20 ns)Description Use this command to set/r

Pagina 79 - :SYSTem:KEY

59Programming ReferenceSCPI Instrument Command ListCommand :PULS:DEL[1]:UNITLong[:SOURce]:PULSe:DELay[1]:UNITForm Set & QueryParameterS | SEC | P

Pagina 80

6 Safety SummarySafety SummaryThe following general safety precautions must be observed during all phases of operation of this instrument. Failure to

Pagina 81 - :SYSTem:PRESet

60 Programming ReferenceSCPI Instrument Command ListCommand :PULS:DOUB[1]:DELLong[:SOURce]:PULSe:DOUBle[1]:DELayForm Set & QueryParameter NumericP

Pagina 82 - :SYSTem:SECurity[:STATe]

61Programming ReferenceSCPI Instrument Command ListCommand :PULS:DOUB[1]:DEL:HOLDLong[:SOURce]:PULSe:DOUBle[1]:DELay:HOLDForm Set & QueryParamete

Pagina 83 - :SYSTem:WARNing[:COUNt]?

62 Programming ReferenceSCPI Instrument Command ListCommand :PULS:HOLD[1]Long[:SOURce]:PULSe:HOLD[1]Form Set & QueryParameterWIDTh | DCYCle | TDEL

Pagina 84 - :TRIGger[:SEQuence[1]]:COUNt

63Programming ReferenceSCPI Instrument Command ListYou cannot set the pulse period if you have selected the CLK-IN connector as the frequency source

Pagina 85 - :TRIGger:IMPedance

64 Programming ReferenceSCPI Instrument Command ListExample To program the pulse width by means of the delay parameters:Command :PULS:TRAN[1]:HOLDLong

Pagina 86 - :TRIGger:SLOPe

65Programming ReferenceSCPI Instrument Command ListCommand :PULS:TRAN[1]:UNITLong[:SOURce]:PULSe:TRANsition[1]:UNITForm Set & QueryParameterS | S

Pagina 87

66 Programming ReferenceSCPI Instrument Command ListCommand :PULS:TRAN[1]:TRALong[:SOURce]:PULSe:TRANsition[1]:TRAilingForm Set & QueryParameter N

Pagina 88

67Programming ReferenceSCPI Instrument Command ListCommand :PULS:TRAN[1]:TRA:AUTOLong:[SOURce]:PULSe:TRANsition[1]:TRAiling:AUTOForm Set & QueryP

Pagina 89 - 3Specifications

68 Programming ReferenceSCPI Instrument Command ListCommand :PULS:WIDT[1]Long[:SOURce]:PULSe:WIDTh[1]Form Set & QueryParameter NumericParameter su

Pagina 90 - Information

69Programming ReferenceSCPI Instrument Command ListExample To set up the external PLL reference:Command :ROSC:EXT:FREQLong[:SOURce]:ROSCillator:EXTer

Pagina 91 - Agilent 81101A Specifications

7Safety SummaryGround the InstrumentTo minimize shock hazard, the instrument chassis and cover must be connected to an electrical protective earth gr

Pagina 92 - Acoustic Noise Emission

70 Programming ReferenceSCPI Instrument Command ListCommand :VOLT[1]Long[:SOURce]:VOLTage[1][:LEVel][:IMMediate][:AMPLitude]Form Set & QueryParame

Pagina 93 - Timing Specifications

71Programming ReferenceSCPI Instrument Command ListCommand :VOLT[1]:OFFSetLong[:SOURce]:VOLTage[1][:LEVel][:IMMediate]:OFFSetForm Set & QueryPara

Pagina 94

72 Programming ReferenceSCPI Instrument Command ListCommand :VOLT[1]:HIGHLong[:SOURce]:VOLTage[1][:LEVel][:IMMediate]:HIGHForm Set & QueryParamete

Pagina 95 - Transition Times

73Programming ReferenceSCPI Instrument Command ListCommand :VOLT[1]:LOWLong[:SOURce]:VOLTage[1][:LEVel][:IMMediate]:LOWForm Set & QueryParameter

Pagina 96 - Level Specifications

74 Programming ReferenceSCPI Instrument Command ListCommand :VOLT[1]:LIMLong[:SOURce]:VOLTage[1]:LIMit[:HIGH]Form Set & QueryParameter NumericPara

Pagina 97 - Clock Sources

75Programming ReferenceSCPI Instrument Command ListCommand :VOLT[1]:LIM:STATLong[:SOURce]:VOLTage[1]:LIMit:STATeForm Set & QueryParameterON | OFF

Pagina 98 - Phase Locked Loop (PLL)

76 Programming ReferenceSCPI Instrument Command ListCommand :STATus:PRESetLong:STATus:PRESetForm Event*RST value Not ApplicableDescription This comman

Pagina 99 - Output Modes

77Programming ReferenceSCPI Instrument Command List1. :STATus:QUEStionable[:EVENt]?2. :STATus:QUEStionable:CONDition?3. :STATus:QUEStionable:ENABle4.

Pagina 100 - Trigger Modes

78 Programming ReferenceSCPI Instrument Command ListCommand :SYST:CHECLong:SYSTem:CHECk[:ALL][:STATe]Form Set & QueryParameterOFF | ON*RST valueON

Pagina 101 - Strobe Output

79Programming ReferenceSCPI Instrument Command ListThe above message is an example of a customized description. Generic descriptions are available in

Pagina 102 - Typical Delays

8 Safety SummarySafety SymbolsCaution (refer to accompanying documents)Protective earth (ground) terminalIn the manuals:WARNINGThe WARNING sign denote

Pagina 103 - Human Interface

80 Programming ReferenceSCPI Instrument Command ListDescription In query form, this command reads the last key pressed. The buffer is emptied by *RST

Pagina 104 - Remote Control

81Programming ReferenceSCPI Instrument Command List • In remote mode only the softkeys below the display and the SHIFT (LOCAL) key are active. Becaus

Pagina 105 - Pulse Parameter Definitions

82 Programming ReferenceSCPI Instrument Command ListCommand :SYST:SECLong:SYSTem:SECurity[:STATe]Form Set & QueryParameterON|OFF*RST valueOFFDescr

Pagina 106 - Pulse Width

83Programming ReferenceSCPI Instrument Command ListCommand :SYST:SETLong:SYSTem:SETForm Set & QueryParameter Block data*RST value Not applicableD

Pagina 107 - Interchannel Delay (Skew)

84 Programming ReferenceSCPI Instrument Command ListCommand :SYST:WARN:STR?Long:SYSTem:WARNing:STRing?Form Query*RST value Not applicableDescription U

Pagina 108 - Linearity

85Programming ReferenceSCPI Instrument Command ListExamples To set up a triggered burst of 16 Single Pulses at Out1, each burst triggered by a positi

Pagina 109 - Pulse Levels

86 Programming ReferenceSCPI Instrument Command ListCommand :TRIG:LEVLong:TRIGger:LEVelForm Set & QueryParameter NumericParameter SuffixV with eng

Pagina 110 - Settling Time

87Programming ReferenceSCPI Instrument Command ListPulse period sources set by :TRIG:SOURPulse period source:TRIG:SOURceinternal oscinternal PLLCLK-I

Pagina 111 - Repeatability

88 Programming ReferenceSCPI Instrument Command List

Pagina 112 - Specifications

8933SpecificationsIn this chapter you will find the specifications of the Agilent 81101A Pulse Generator.At the end of this chapter, “Pulse Parameter

Pagina 113

9About this BookAbout this BookThis guide provides reference information primarily for programming the Agilent 81101A via remote control.Chapter 1 “G

Pagina 114

90 SpecificationsDeclaration of ConformityDeclaration of ConformityManufacturerAgilent Technologies Deutschland GmbHBoeblingen Verification SolutionsH

Pagina 115

91SpecificationsAgilent 81101A SpecificationsAgilent 81101A SpecificationsGeneralEnvironmental ConditionsSafety IEC1010, CSA1010Power requirements

Pagina 116

92 SpecificationsAgilent 81101A SpecificationsWeight Net8.5 kg Single Channel9.2 kg Dual ChannelShipping13.8 kg Dual ChannelRecalibration period 1

Pagina 117 - Front Panel Controls

93SpecificationsAgilent 81101A SpecificationsTiming SpecificationsPeriod Period can also be entered as frequency. There are 2 period generation sourc

Pagina 118 - 81101-91021

94 SpecificationsAgilent 81101A SpecificationsWidthCan be entered as absolute width, duty cycle or trailing edge delay.Duty cycle values from 0.1% to

Pagina 119

95SpecificationsAgilent 81101A SpecificationsDouble Pulse DelayDouble pulse delay and delay are mutually exclusive. Double Pulse delay is the delay b

Pagina 120

96 SpecificationsAgilent 81101A SpecificationsLeading and trailing edges can be programmed independently within the following ranges (Maximum ratio 1:

Pagina 121

97SpecificationsAgilent 81101A SpecificationsLevel ParametersLevel parameters can be entered as high/low level in terms of voltage or current or offs

Pagina 122

98 SpecificationsAgilent 81101A SpecificationsClock / PLL Reference InputRear panel BNC connector used as: • External system clock input: pulse freque

Pagina 123 - DECLARATION OF CONFORMITY

99SpecificationsAgilent 81101A SpecificationsExternal Clock • The output period is determined by the signal at clock input. Frequency accuracy can be