Agilent Technologies E1356A Manuale di Servizio Pagina 30

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 68
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 29
Test 2-2: Leakage
Test
The test verifies the input impedance by measuring the voltage drop across a
known resistor in series with the input impedance. Leakage is measured
from HI to LO, HI to Chassis, and LO to Chassis. Because of the solid state
nature of the switches and input protection, the leakage is measured at both
+10 Vdc and -10 Vdc.
HI to LO Leakage 1. Make Hardware Connections
Turn power supply, DMM, and mainframe power OFF
Connect DMM, power supply, and resistor as shown in Figure
2-6
Turn power supply, DMM, and mainframe power ON
Set power supply output to +10 Vdc ± 0.1 Vdc
2. Check Direct Terminals Leakage
Send *RST to FET Multiplexer
Send TRIG SGL to DMM
Record the DMM reading in Table 2-1
Positive Polarity, HI to LO, Direct
The DMM measurement should be less than 0.010 Vdc. A
measurement out of this range indicates a failure of the FET
Multiplexer and troubleshooting/repair/replacement procedures,
described in Chapter 4, should be performed before proceeding
with Test 2-2
Agilent E1300B/E1301B
Figure 2-6. Positive HI to LO Leakage Connections
2-12 Verification Tests
Vedere la pagina 29
1 2 ... 25 26 27 28 29 30 31 32 33 34 35 ... 67 68

Commenti su questo manuale

Nessun commento