Specifi cation condition
Specification and supplemental
characteristics are defined at 23° C ±
5° C < 60% relative humidity (R.H.)
The supplemental characteristics
entries in the following specifica-
tion are not warranted but they
provided useful information about the
functions and performances of the
instruments.
Agilent B2200A
fA leakage switch
mainframe (back
panel)
Supplemental characteristics (B2200A/B2210A)*
1
Offset current 10 fA*
2
I-V port
IM noise (RMS) 0.6 fA*
3
I-V port
Channel crosstalk capacitance < 1 pF/ch I-V port
< 3 pF/ch AUX port
Offset voltage < 50 µV I-V port
< 80 µV AUX port
Settling time*
4
2.0 sec at 50 fA
Band width (at –3 dB) 30 MHz AUX port
Guard capacitance < 145 pF*
5
I-V port
Additional C measurement error < ± 1 % + 0.2 pF*
6
AUX port
Supplemental characteristics (B2201A/B2211A)*
1
Offset current 100 fA*
7
I-V port
IM Noise (RMS) 5 fA*
3
I-V port
Channel crosstalk capacitance < 0.5 pF/ch I-V port
< 3 pF/ch AUX port
Offset voltage < 80 µV I-V port
< 100 µV AUX port
Settling time*
4
2.0 sec at 300 fA
Band width (at –3 dB) 30 MHz AUX port
Guard capacitance < 145 pF*
5
I-V port
Additional C measurement error < ± 1 % + 0.2 pF*
6
AUX port
*1. B2200A only supports B2210A fA leakage switch module.
B2201A only supports B2211 14ch low leakage switch module.
Mixed confi guration is not supported.
*2. The offset current when zero volts is applied to all input and output channels.
This measurement is made on a port, 5 seconds after a switching a relay.
*3. Measured 100 PLC by 4156C when zero volts are applied to all other paths.
*4. After 10 V applied.
*5. The guard capacitance of the closed port on input and output ports when
4 modules per frame are installed.
*6. The additional error using the C-compensation algorithm for Agilent 4284 at
1 kHz to 1 MHz, <1000 pF and 3 m cable.
*7. The offset current when zero volts is applied to all input and out channels.
This measurement is made on a port, 60 seconds after a switching a relay.”
Accessories & cables
Agilent 16443A Light pen for B2200A/
B2201A
Agilent 16494A Triaxial cable
Agilent 16493K Kelvin triaxial cable
(for input port)
(Between 4155/56
series or E5270 series
and B2200/2201A
input port)
Agilent 16494B Kelvin triaxial cable
(for output port)
Agilent 16494F CMU input cable
Agilent 16493N GND cable
(Between GND of
E5270/41501 and
B2201A/B2200A)