Agilent E361xA 60W BENCH SERIES DC POWER SUPPLIESOPERATING AND SERVICE MANUAL FOR MODELS:Agilent E3614A, Serials KR83503035 and aboveAgilent E3615A, S
1-10ply above 100% rated output. Limit your programming voltage to 10 Vdc.Remote Programming Connections. Remote programmingrequires changing settings
1-11Setting Voltage and Current. Turn the slave unit's CURRENTcontrol fully clockwise. Adjust the master unit's controls to set thedesired o
1-12AUTO-SERIES OPERATIONAuto-series operation permits equal or proportional voltagesharing, and allows control of output voltage from one masterunit.
1-13Remote Sensing. To remote sense with auto-series operation,set SENSE switch of the master unit and set SENSE switch of theslave unit to remote.Rem
1-14LOAD CONSIDERATIONSThis section provides information on operating your supply withvarious types of loads connected to its output.PULSE LOADINGThe
A-1SERVICE INFORMATIONFigure A-1. Block DiagramPRINCIPLES OF OPERATION(Block Diagram Overview)Throughout this discussion, refer to both the block dia
A-2series pass transistors. Whenever the output voltage is belowthe sloping V1 line, the control circuit inhibits four SCRs andthe input capacitors ch
A-3a.Perform turn-on checkout procedure given in page 1-7.b.Perform the CV and CC Load Regulation perfor-mance tests given in the following paragraphs
A-4CONSTANT VOLTAGE (CV) TESTSCV Setup. For all CV tests set the output current at full ratedoutput to assure CV operation. The onset of constant curr
A-5Figure A-5. Load Transient Response Time WaveformPARD(Ripple and Noise)Definition: PARD is the Periodic and Random Deviation ofthe dc output voltag
1-2SAFETY SUMMARYThe following general safety precautions must be observed during all phases of operation, service, and repair of this instrument. Fai
A-6a. Connect the test equipment as shown in Figure A-7.b. Turn the supply's power on and turn CURRENT con-trol fully clockwise.c. Turn up output
A-7e. Record output voltage across Rs and convert it to current by dividing this voltage by Rs.f. Adjust autotransformer to the high line voltage (127
A-8Ammeter and CC Set CalibrationTo calibrate ammeter and CC set, proceed as follows:a. Connect test setup on Figure A-9.b. Turn VOLTAGE and CURRENT c
A-9After the trouble has been isolated to one of the feedbackloops, troubleshooting can proceed as described in Tables A-4, A-5, or A-6.Series Regulat
A-10SYMPTOM CHECKS AND PROBABLE CAUSESPoor Load Regulation(Constant Voltage)a. Refer to "Measurements Techniques" paragraph.b. Check +10 V
A-11STEP ACTION RESPONSE PROBABLE CAUSE4 Check voltage from pin 13 to pin 12 of U9.a. Measured voltage is positive.b. Measured voltage is negative.a.
A-12REPLACEABLE PARTSINTRODUCTIONThis section contains information for ordering replacementparts. Table A-10 lists parts by reference designator
Table A-10. Replaceable Parts ListReference DesignatorNumberQ'ty Description Model Mfr. P/N Mfr.CodeE3614A 1 60W BENCH POWER SUPPLY - E3614A M
Table A-10. Replaceable Parts List (Cont'd)Reference Designator Agilent PartNumberQ'ty Description Model Mfr. P/N Mfr.CodeC18,21,24,27 01
Table A-10. Replaceable Parts List (Cont'd)Reference Designator Agilent PartNumberQ'ty Description Model Mfr. P/N Mfr.CodeR24,26,27,37,38
1-3Table of ContentsSAFETY SUMMARY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
R62 0698-4514 1 RESISTOR 105K +-1% .125W TF TC=0+-100 15 28480R62 0757-0481 1 RESISTOR 475K +-1% .125W TF TC=0+-100 16 28480Table A-10. Replaceable
R109 0698-3153 1 RESISTOR 3.83K +-1% .125W TF TC=0+-100 ALL 28480R115 0757-0462 1 RESISTOR 75K +-1% .125W TF TC=0+-100 14 28480Table A-10. Replacea
Table A-10. Replaceable Parts List (Cont'd)Reference Designator Agilent PartNumberQ'ty Description Model Mfr. P/N Mfr.CodeCR21,22,23,24,2
Table A-11. Component ValueComponent ModelE3614A E3615A E3616A E3617AC2 1000UF 25V +-20% AL-ELECTLT 470UF 50V +-20% AL-ELECTLT 330UF 50V +-20% AL-EL
Manual SupplementSupplement Agilent Part Number : 5959-5336, Edition 4Supplement Print Date : 14 April, 2000This supplement updates the following docu
1-10-1Voltage and Current Programming of the E3614A/15A/16A/17A with a Voltage and Current SourceRemote analog voltage programming permits control of
1-10-2Programming Voltage Common to the Minus OutputFigure 2Set the CV switch down on the rear panel, and all others up.Vin = 1/A x VoutVout = A x Vin
1-10-3Programming Voltage Common to the Minus OutputFigure 4The output will always be the same or less than the programming voltage.The M/S2 switch mu
1-10-4Constant Current ModeThe E3614A/15A/16A/17A may be programmed for constant current with an analog voltage or current. Constant current with anal
1-10-5Figure 6Set the CC switch down, and all others up.Iin = 1/A x IoutIout = A x VinWhere Iout is the power supply output current in amps.Iin is the
1-4GENERAL INFORMATIONINTRODUCTIONThis manual describes all models in the Agilent E361xA 60WBench Power Supply family and unless stated otherwise, the
1-10-6Voltage and Current Programming of the E3614A/15A/16A/17A with ResistorsRemote programming with resistors permits control of the regulated outpu
1-10-7Remote Resistor Programming, Constant CurrentFigure 8Set the CC switch down on the rear panel, and all others up.Iout = A x [VREF x {R/(R + R2 +
CERTIFICATIONAgilent Technologies certifies that this product met its published specifications at time of shipment from the factory. AgilentTechnologi
DECLARATION OF CONFORMITYaccording to ISO/IEC Guide 22 and EN 45014Manufacturer’s Name: Agilent Technologies, Inc.Manufacturer’s Address: 345-15, Kasa
1-5be grounded or the power supply can be operated float-ing at up to a maximum of 240 Volts off ground. Total out-put voltage to ground must not exce
1-6Table 1. Specifications and Operating Characteristics (Cont’d)*REMOTE PROGRAMMING SPEEDMaximum time required for output voltage to change from init
1-7e.Replace the cover and mark the supply clearly with a tag or label indicating the correct line voltage and fuse that is in use. Figure 2. Line Vol
1-8that the current limit value can be set from zero to maximum rated value.OPERATING MODESThe setting of the rear panel switch determines the operati
1-9REMOTE OPERATING MODES Remote operating modes discussed below are remote voltagesensing and remote voltage programming. You can set up the unitfor