Performance
Capturedepth 4GBmemory,512MSaofIQdata
Frequencyandtimespecifications
Frequency range
AllRFports(optionsareperTRXmodule)
OptionE6640A-504
OptionE6640A-5WC
OptionE6640A-506
380MHzto3.8GHz
1.1to1.8GHz,2.3to2.6GHz,and4.8to6.0GHz
380MHzto6.0GHz
Specified frequency range (dependent on selected frequency range option)
380to495MHz
695to920MHz
1425to1485MHz
1620to2030MHz
2300to2700MHz
3400to3800MHz
4900to6000MHz
Frequency reference
Accuracy,agingrate,stability RefertoTimebaseSpecifications
CW measurement frequency accuracy
Accuracy
(Transmitterfrequencyxfrequencyreferenceaccuracy)±50Hz)typical
Resolution
1Hztypical
Analysis bandwidth
Maximumbandwidth
E6640A-B40
E6640A-B85
380to495MHz
695to800MHz
Allotherspecifiedfrequencyranges
Upto40MHzanalysisbandwidth
Upto40MHz
Upto60MHz
Upto80MHz
E6640A-B1X
380to495MHz
695to800MHz
3400to3800MHz
Allspecifiedfrequencyranges
Upto40MHz
Upto60MHz
Upto100MHz
Upto160MHz
Triggering
Trigger
Sequenceanalyzer
IQanalyzer
Freerun,external1,external2,RFburst,video,internal
Freerun,external1,external2,RFburst,video,line,periodic
Triggerdelayrange –15to500ms
Resolution 0.1μs
Amplitudeaccuracyandrangespecifications
Input level ranges (average power)
RF3I|OandRF4I|O(halfduplex) –70to+30dBm
RFIO1andRFIO2(fullduplex) –65to+36dBm