17. 20/30 Loop Phase Lock Adjustments
19. Set the RF Section LINE switch to STANDBY. Disconnect the DVM
from
AlOA4TP3
and the second spectrum analyzer from
AlOA4J2
OUT 160-166 MHz.
20. Reinstall
AlOA4
PLL3 Up Converter Assembly in the RF Section.
Reconnect the cable 7 (violet) to AlOA5J4 SCAN
5.1
MHz OUT,
the cable 6 (blue) to
AlOA4J2
OUT 160-166 MHz, and the cable 8
(gray) to AlOA4J3 IN 100 MHz.
Adjustments 3-159
Vedere la pagina 298