Agilent Technologies InfiniiVision 3000 X-Series Manuale Utente Pagina 19

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 32
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 18
19
DSOX3000 Series (digital signal oscilloscope)
MSOX3000 Series (mixed signal oscilloscope)
DSOX3012A
MSOX3012A
DSOX3014A
MSOX3014A
DSOX3024A
MSOX3024A
DSOX3032A
MSOX3032A
DSOX3034A
MSOX3034A
DSOX3052A
MSOX3052A
DSOX3054A
MSOX3054A
Characteristics
Analog
bandwidth* (-3dB)
100 MHz 100 MHz 200 MHz 350 MHz 350 MHz 500 MHz 500 MHz
Analog input
channels
2442424
Calculated rise
time
≤ 3.5 ns ≤ 3.5 ns ≤ 1.75 ns ≤ 1 ns ≤ 1 ns ≤ 800 ps ≤ 800 ps
All 3000 X-Series models
Hardware bandwidth limits 20 MHz selectable
Input coupling AC, DC
Input impedance Selectable: 1 MΩ ± 1%; II 14 pF, 50 Ω ± 1.5%
Input sensitivity range 1 mV/div to 5 V/div**
Maximum sample rate 2 GSa/s per channel, 4 GSa/s half-channel interleaved
Maximum memory depth
(record length)
1 Mpt per channel, 2 Mpts half-channel interleaved (standard),
2 Mpts per channel, 4 Mpts half-channel interleaved (optional with DSOX3MEMUP)
Display 8.5-inch WVGA with 64 levels of intensity grading
Waveform update rate (max) 1,000,000 waveforms/s
Vertical resolution 8 bits
Horizontal resolution 2.5 ps
Maximum input voltage CAT I 300 Vrms, 400 Vpk; transient overvoltage 1.6 kVpk
CAT II 300 Vrms, 400 Vpk
With N2862A, N2863A or N2890A 10:1 probe: 300 Vrms
DC vertical accuracy ±[DC vertical gain accuracy + DC vertical offset accuracy + 0.25% full scale] **
DC vertical gain accuracy*
±2% full scale**
Channel-to-channel isolation > 100:1 from DC to maximum specified bandwidth of each model
(measured with same V/div and coupling on channels)
Offset range
±2 V (1 mV/div to 200 mV/div)
±50 V (> 200 mV/div to 5 V/div)
DC vertical offset accuracy ±0.1div ± 2mV ± 1% of offset setting
Oscilloscopes redefined: Breakthrough technology delivers more scope for the same budget
Performance characteristics
* Denotes warranted specifications, all others are typical.
Specifications are valid after a 30-minute warm-up period and ±10 °C from firmware calibration temperature.
** 1 mV/div and 2 mV/div are a magnification of 4 mV/div setting. For vertical accuracy calculations, use full scale of 32 mV for 1 mV/div and
2 mV/div sensitivity setting.