Performance characteristics (continued)
Trigger system (continued)
Selections Edge,pulsewidth,pattern,TV,duration
Edge Triggeronarising,falling,oralternatingedgeofanysource
Pulsewidth Triggerwhenapositive-ornegative-goingpulseislessthan,greaterthan,orwithinaspecifiedrange
onanyofthesourcechannels.
Minimumpulsewidthsetting:
5ns(DSO501xA)
2ns(DSO503xA,DSO505xA)
Maximumpulsewidthsetting:10s
Pattern Trigger at the beginning of a patternofhigh,low,anddon'tcarelevelsand/orarisingorfallingedge
establishedacrossanyofthechannels,butonlyafterapatternhasbeenestablishedforaminimumof
2nsec.Thechannel’shighorlowlevelisdefinedbythatchannel’striggerlevel.
Sequence ArmonEventA,triggeronEventB(edgeorpattern),withoptiontoresetonEventCortimedelay
TV Triggerusinganyoscilloscopechannelonmostanalogprogressiveandinterlacedvideostandards
includingHDTV/EDTV,NTSC,PAL,PAL-MorSECAMbroadcaststandards.Selecteitherpositiveor
negativesyncpulsepolarity.ModessupportedincludeField1,Field2,allfields,alllines,oranyline
withinafield.TVtriggersensitivity:0.5divisionofsyncsignal.Triggerholdofftimecanbeadjustedin
half field increments.
Duration Triggeronamulti-channelpatternwhosetimedurationislessthanavalue,greaterthanavalue,
greaterthanatimevaluewithatimeout,orinsideoroutsideofasetoftimevalues.
Minimumdurationsetting:2ns
Maximumdurationsetting:10s
CAN TriggeronCAN(ControllerAreaNetwork)version2.0Aand2.0Bsignals.Triggeronthestartofframe
(SOF)bit(standard).N5424AoptionsupportstriggeringonremoteframeID(RTR),dataframeID
(~RTR),remoteordataframeID,dataframeIDanddata,errorframe,allerrors,acknowledgeerrorand
overloadframe.
FlexRay Trigger on FlexRayFrames,errors,eventsandcycle-multiplexedtriggering.N5432Coroption
FLXsupportsalsotriggeringonparticularframetypessymbolically,suchasStartupframes,Nullframe,
Syncframe,etc.,aswellasBooleanNOTframetypes.
LIN Trigger on LIN (Local Interconnect Network) sync break at beginning of message frame (standard).
N5424AoptionsupportstriggeringonframeID.
I
2
C Trigger on I
2
C (Inter-ICbus)serialprotocolatastart/stopconditionoruserdefinedframewithaddress
and/ordatavalues.Alsotriggeronmissingacknowledge,addresswithnoacq,restart,EEPROMread,
and 10‑bit write.
SPI Trigger on SPI (SerialProtocolInterface)datapatternduringaspecificframingperiod.Supportspositive
andnegativeChipSelectframingaswellasclockIdleframinganduser-specifiednumberofbitsper
frame.
RS-232/UART View the informationsentoveraRS-232serialbus.Displayreal-timetime-aligneddecodeoftransmit
andreceivelines.TheapplicationalsoenablestriggeringonRS-232/UARTconditions.
MIL‑STD 1553 Trigger on specific Command/StatusWords,DataWords,anderrorconditions.
AutoScale Findsanddisplaysallactivechannels,setsedgetriggermodeonhighest-numberedchannel,sets
verticalsensitivityonchannels,timebasetodisplay~1.8periods.Requiresminimumvoltage>
10mVpp,0.5%dutycycleandminimumfrequency>50Hz.