Agilent Technologies N5230C Guida Utente Pagina 58

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 72
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 57
52 User’s Guide
U3025AE10 Operational Check
Operational Check Procedure for E8363/4C and N5230C Systems
1. Connect the ECal or the mechanical cal kit to Port 1 or the port you are testing. Torque
to 8 in-lb. For more information press the Help button, see Figure 50.
2. Perform a 1-Port Calibration on Port 1. On the PNA, press Response > CAL > Start Cal
> Calibration Wizard.
a. If using a mechanical cal kit, select SmartCal.
b. If using an ECal module, select ECal.
3. Continue following the Cal Wizard prompts. In the “Select Calibration Ports and ECal
Module” window, press Clear All and select Port 1, then press Next.
Figure 50 1-Port Calibration
4. Continue to follow the prompts until the “Calibration Completed” window appears.
Vedere la pagina 57
1 2 ... 53 54 55 56 57 58 59 60 61 62 63 ... 71 72

Commenti su questo manuale

Nessun commento