User’s Guide U3025-90001 9
U3025AE10 General Specifications
Typical Reflection Tracking
Specifications for the Test Set are typical. System performance for the PNA and Test Set
are only characteristic and are intended as non-warranted information.
NOTE Typical specifications are based on 1 to 2 units performance, refer to Table 4.
Table 4 Typical Reflection Tracking Loss
1,
2
1. Reflection tracking loss is generally source path loss + receiver path loss.
2. Does not include margin of ± 2 dB, or PNA accuracy ± 1.5 dB up to 40 GHz
(± 2 dB 40 to 50 GHz).
Frequency
Standard (dB)
(typical)
Option 001 (dB)
(typical)
10 MHz to 45 MHz +3.0 +2.5
45 MHz to 500 MHz +6.5 +6.0
500 MHz to 2 GHz +4.0 +3.5
2 GHz to 10 GHz +2.0 +1.5
10 GHz to 20 GHz 0 0.5
20 GHz to 30 GHz 3.0 8.0
30 GHz to 40 GHz 4.5 9.0
40 GHz to 45 GHz 7.0 10.0
45 GHz to 50 GHz 8.5 12.0