Agilent Technologies U4154A Guida di Installazione Pagina 55

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
Vedere la pagina 54
Installing Software Components 3
U4154A Logic Analyzer Installation Guide 55
To upgrade the module FPGA using this tool
1 Access the upgrade tool by clicking <Program Files>\
Agilent Technologies\Logic Analyzer\
agAmpAxieUpgrade.exe.
2 Select the Upgrade Type as Blade Interface FPGA.
The expected versions of FPGA are displayed for each of
the modules.
3 From the Instance listbox, select the module which
requires the FPGA upgrade. The modules are listed by the
slot numbers in which these are installed in the AXIe
chassis. You can use the slot number indicated in the
error message displayed at the start to find out the
Vedere la pagina 54
1 2 ... 50 51 52 53 54 55 56 57 58 59 60 ... 67 68

Commenti su questo manuale

Nessun commento