44 Understanding the Agilent E1330B Digital I/O Module
Chapter 4
Using the Handshake Modes
Handshaking ensures correct transfer of data between devices. You must set
both the mode and the timing to establish correct handshaking. SCPI
commands support the following modes of handshaking:
-- LEADing Edge
-- TRAiling Edge
-- PULSe
-- PARTial
-- STRobe
-- NONE
These SCPI commands set the type of handshake mode used:
[SOURce:]DIGital:DATAn[:type]:HANDshake[:MODE] <mode>
[SOURce:]DIGital:HANDshake
n[:MODE] <mode>
These SCPI commands set the timing of the handshake (where timing
applies):
[SOURce:]DIGital:DATAn[:type]:HANDshake:DELay <time>
[SOURce:]DIGital:HANDshake
n:DELay <time>