E5062A Performance Test Record (for Option 150/250)
Direction: S12 (Option 250 only)
For Serial Prefix : MY442 and above
Direction: S21
Direction: S12 (Option 250 only)
Frequency range [Hz] IF bandwidth [Hz] Test limit [dB] Test result [dB]
300 k - 100 M 10 -115
100 M - 1.5 G 10 -115
1.5 G - 3 G 10 -115
300 k - 100 M 3 k -90
100 M - 1.5 G 3 k -90
1.5 G - 3 G 3 k -90
Frequency range [Hz] IF bandwidth [Hz] Test limit [dB] Test result [dB]
300 k - 1 M 10 -115
1 M - 100 M 10 -120
100 M - 1.5 G 10 -120
1.5 G - 3 G 10 -120
300 k - 1 M 3 k -90
1 M - 100 M 3 k -95
100 M - 1.5 G 3 k -95
1.5 G - 3 G 3 k -95
Frequency range [Hz] IF bandwidth [Hz] Test limit [dB] Test result [dB]
300 k - 1 M 10 -115
1 M - 100 M 10 -120
100 M - 1.5 G 10 -120
1.5 G - 3 G 10 -120
300 k - 1 M 3 k -90
1 M - 100 M 3 k -95
100 M - 1.5 G 3 k -95
1.5 G - 3 G 3 k -95