Commands by Subsystem 5
Agilent InfiniiVision 6000 Series Oscilloscopes Programmer's Guide 521
:TRIGger:FLEXray:ERRor:TYPE
(see page 798)
Command Syntax
:TRIGger:FLEXray:ERRor:TYPE <error_type>
<error_type> ::= {ALL | HCRC | FCRC}
Selects the FlexRay error type to trigger on. The error type setting is only
valid when the FlexRay trigger mode is set to ERRor.
ALL — triggers on ALL errors.
HCRC — triggers on only Header CRC errors.
FCRC — triggers on only Frame CRC errors.
Query Syntax
:TRIGger:FLEXray:ERRor:TYPE?
The :TRIGger:FLEXray:ERRor:TYPE? query returns the currently selected
FLEXray error type.
Return Format
<error_type><NL>
<error_type> ::= {ALL | HCRC | FCRC}
See Also "Introduction to :TRIGger Commands" on page 479
":TRIGger:FLEXray:TRIGger" on page 528
NOTE
This command is only valid when the FLEXray triggering and serial decode option (Option
FLX) has been licensed.