Agilent Technologies InfiniiVision 6000 Series Manuale di Servizio Pagina 711

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 934
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 710
Obsolete and Discontinued Commands 7
Agilent InfiniiVision 6000 Series Oscilloscopes Programmer's Guide 711
:DISPlay:CONNect
(see page 798)
Command Syntax
:DISPlay:CONNect <connect>
<connect> ::= {{ 1 | ON} | {0 | OFF}}
The :DISPlay:CONNect command turns vectors on and off. When vectors
are turned on, the oscilloscope displays lines connecting sampled data
points. When vectors are turned off, only the sampled data is displayed.
Query Syntax
:DISPlay:CONNect?
The :DISPlay:CONNect? query returns the current state of the vectors
setting.
Return Format
<connect><NL>
<connect> ::= {1 | 0}
See Also ":DISPlay:VECTors" on page 265
NOTE
The :DISPlay:CONNEct command is an obsolete command provided for compatibility to
previous oscilloscopes. Use the :DISPlay:VECTors command (see page 265) instead.