2 1100 Series Capillary LC System Manual
1 Installing your Capillary LC System
Site Requirements
A suitable environment is important to ensure optimum performance of the
Capillary LC system.
Power Consideration
The modules power supply has wide ranging capability (see Table 1 on page 4).
It accepts any line voltage in the range described in the above mentioned table.
Consequently there is no voltage selector in the rear of the modules. There are
also no externally accessible fuses, because automatic electronic fuses are
implemented in the power supply.
Power Cords
Different power cords are offered as options with the modules. The female end
of each of the power cords is identical. It plugs into the power-input socket at
the rear of the instruments. The male end of each of the power cords is
different and designed to match the wall socket of a particular country or
region.
To disconnect the modules from line, unplug the power cord. The power supply still
uses some power, even if the power switch on the front panel is turned off.
Shock hazard or damage of your instrumentation can result, if the devices are
connected to a line voltage higher than specified.
Never operate your instrumentation from a power outlet that has no ground
connection. Never use a power cord other than the power cord designed for your
region.
Never use cables other than the ones supplied by Agilent Technologies to ensure
proper functionality and compliance with safety or EMC regulations.