Agilent Technologies 1100 Series Manuale delle Istruzioni Pagina 171

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
Vedere la pagina 170
1100 Series Capillary LC System Manual 161
Options 6
Replacing or Cleaning Parts
CAUTION
The quartz block can be cleaned with alcohol. DO NOT touch the inlet and outlet
windows at the quartz block.
1 Disconnect the capillaries from the capillary holder
and remove the flow cell.
2 Unscrew the cell body from the holder.
3 Unscrew the capillaries from the flow cell. DO NOT
use the adapter at this time!
4 Using a toothpick, press on the plastic part and slide
the quartz body out of the cell housing.
Vedere la pagina 170
1 2 ... 166 167 168 169 170 171 172 173 174 175 176 ... 195 196

Commenti su questo manuale

Nessun commento