112 Agilent E6474A User’s Guide
4 Use Your System
Test 2 - Receiver Testing by Measuring a Known Channel
Preparing the Hardware
1 Connect the system as shown in Figure 48.
2 Turn on the Agilent receiver.
Figure 48 System configuration for measuring a known channel in the system
NOTE
The Agilent receiver requires a 30-minute warm-up period to meet the
guaranteed specifications. There is a three-minute delay after power-up or
re-configuration, before the internal temperature sensor is activated and
measurements are nominally correct.