Agilent PXT Wireless Communications Test Set
User’s Guide
114
Connected DRX On/Off
Controls the drx-Config element of the RRC Connection Reconfiguration message to be set to setup or
release. This enables or disables Connected Mode DRX in the UE.
This setting overwrites the value defined in the currently loaded scenario file, when the
BSE emulator is not running (Mode > BSE > Emulator Mode = Stop
is selected).
Although the RRC Reconfiguration message can be overwritten in both FDD and TDD,
the behavior is different when it is configured. In FDD, the Connected Mode DRX
configuration is observed and data is only sent to the UE when it is listening to the
downlink. In TDD, data may be sent to the UE at any time.
Overwrites Scenario File Value
BSE > Mode Setup> More > RRC Settings > More
longDRX-Cycle
Enables you change the setting of the longDRx-cycle that is present in the RRC Connection Reconfiguration
message of the currently loaded scenario file.
Overwrites Scenario File Value
Range
SF10 | SF20 | SF32 | SF40 | SF64 | SF80 | SF128 | SF160 | SF256 | SF320
| SF512 | SF640 | SF1024 | SF1280 | SF2048 | SF2560
Dependencies and/or Couplings
Connected DRx On/Off must be set to On.
BSE > Mode Setup> More > RRC Settings > More
DRX-StartOffset
Enables you change the value of the longDRX-CycleStartOffset that is present in the RRC Connection
Reconfiguration message of the currently loaded scenario file. This determines where in the longDRX-Cycle,
the onDurationTimer starts.
Overwrites Scenario File Value
Dependencies and/or Couplings
Connected DRx On/Off must be set to On.
BSE > Mode Setup> More > RRC Settings > More