10 Adjust the channel 1 high-frequency pulse response for an %
overshhot of 5% mininum to 7% maximum.
If a low loss 50 cable is used (such as 8120-4949), the % overshoot should be
6% minimum to 8% maximum.)
11
Repeat steps 2 through 10 substituting Channel 2 for Channel 1 and
V(2) for V(1).
High-frequency pulse response adjustment locations
High-frequency pulse
response adjustments
Figure 4-3
Service
Adjusting the Oscilloscope
4–28