Agilent Technologies 54657A Specifiche Pagina 89

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 188
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 88
5
Set the time base to 200 µs/div, then center the signal on the display
with the delay knob (delay about 800 µs).
6 Press Main/Delayed , then press the Delayed softkey.
7
Set the delayed sweep to 20 µs/div, then set the expanded portion
over the VITS (delay about 988.8 µs).
Second VITS Displayed
This figure shows the
second vertical
interval test signals
displayed with delayed
sweep.
Figure 3-10
Using Option 005 Enhanced TV/Video Trigger
To analyze video waveforms with Option 005
3–17