Agilent Technologies E4438C Guida di Installazione Pagina 106

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
Vedere la pagina 105
Troubleshooting
Troubleshooting Performance Test Problems
1-94
AM Frequency Response •Perform:
Internal AM Accuracy (This is part of the Internal AM
Accuracy and Distortion performance test.)
If still failing and:
problem is at carrier frequencies 4GHz,
troubleshoot the following assemblies:
A13 Output
problem is only at carrier frequencies > 4GHz,
troubleshoot the following assemblies:
A14 Extended Frequency Output (Option 506)
FM Frequency Response Perform:
Internal FM Accuracy (This is part of the Internal FM
Accuracy and Distortion performance test.)
If still failing, troubleshoot the following assemblies:
A17 Synthesizer (Option 501, 502, 503, 504)
or A16 Frac-N (Option UNJ or Option 506)
Phase Modulation Frequency Response Perform:
Internal PM Accuracy (This is part of the Phase
Modulation Accuracy and Distortion performance
test.)
If still failing, troubleshoot the following assemblies:
A17 Synthesizer (Option 501, 502, 503, 504)
or A16 Frac-N (Option UNJ or Option 506)
DCFM Frequency Offset Relative to CW Perform:
DCFM Calibration
If still failing, troubleshoot the following assemblies:
A18 Reference
A23 Motherboard
Table 1-40 Troubleshooting Performance Test Problems
Performance test that is failing... Action to perform...