Agilent Technologies E8247C and E8257C/D Option H30User’s and Service Guide Supplement
2 User’s and Service Guide SupplementDescription The Option H30 adds a mixer and coax switch to enable the E8247C or the E8257C/D to upconvert a RF si
User’s and Service Guide Supplement 3 Nominal Performance Characteristics Nominal values indicate expected performance or describe product performa
4 User’s and Service Guide SupplementFigure 1 Upconverted RF FlatnessFigure 2 Upconverted RF Conversion Loss0510152000.40.811.41.622.42.633.23.63.844.
User’s and Service Guide Supplement 5 Operation and Performance VerificationDocumentation Conventions The following key conventions are used throug
6 User’s and Service Guide SupplementPerformance VerificationThe E8247C and E8257C/D Option H30 functionality can be verified using the following equi
User’s and Service Guide Supplement 7 TroubleshootingIf there is no signal a problem has occurred. First check the equipment setup, verifying outpu
8 User’s and Service Guide SupplementUpconverter MixerThe following procedure will aid in determining if the upconverter mixer is functional:CAUTIONPo
User’s and Service Guide Supplement 9 Repair InformationFigure 4 Top ViewFigure 5 Semi-rigid Cables (front view)W102W104 W101A9 MID Board W103W105W
10 User’s and Service Guide SupplementFigure 6 Ribbon CableFigure 7 Bracket (rear panel)A100MP101W106A9MP107MP108NutWasher
User’s and Service Guide Supplement 11 Replaceable PartsReference DesignatorDescription AgilentPart NumberQtyA100 Mixer 6 GHz to 46 GHz (Opt H30 wi
12 User’s and Service Guide SupplementSafety and Regulatory InformationIntroductionReview this product and related documentation to familiarize yourse
User’s and Service Guide Supplement 13 Compliance with German Noise Requirements This is to declare that this instrument is in conformance with the
14 User’s and Service Guide SupplementWarningsWARNINGThe WARNING notice denotes a hazard. It calls attention to a procedure, practice, or the like, wh
User’s and Service Guide Supplement 15 CautionsCAUTIONThe CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice,
16 User’s and Service Guide SupplementInstrument MarkingsWhen you see this symbol on your instrument, you should refer to the instrument’s instruction
User’s and Service Guide Supplement 17 Contacting AgilentBy internet, phone, or fax, get assistance with all your test and measurement needs.This i
18 User’s and Service Guide Supplement
Agilent Technologies E8247C and E8257C/D Option H30User’s and Service Guide SupplementUse this manual with the following documents:PSG Family User’s G
ii Warranty StatementTHE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED “AS IS,” AND IS SUBJECT TO BEING CHANGED, WITHOUT NOTICE, IN FUTURE EDITIONS.
iiiSafety Notes The following safety notes are used throughout this document. Familiarize yourself with each of these notes and its meaning before per
ContentsContents-1E8247C and E8257C\D Option H30Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .