13
Interfaces GPIB (IEEE-488.2,1987) with listen and talk, RS-232,
and 10BaseT LAN interface.
Control languages SCPI version 1997.0. Also will emulate most applicable
Agilent 836xxB, Agilent 837xxB, and Agilent 8340/41B
commands, providing general compatibility with ATE
systems which include these signal generators.
IEEE-488 functions SH1, AH1, T6, TE0, L4, LE0, SR1, RL1, PP0, DC1,
DT0, C0, E2.
ISO compliant This family of signal generators is manufactured in an
ISO-9001 registered facility in concurrence with Agilent
Technologies commitment to quality.
Power requirements 90 to 132 Vac 50 to 60 Hz, or 195 to 267
Vac 50 to 60 or 400 Hz, (automatically selected),
300 W maximum.
Operating temperature range 0 to 55 °C
Storage temperature range
1
–40 to 71 °C
Shock and vibration
Operating random vibration 5 to 500 Hz, 0.21 g rms
Survival swept sine vibration 5 to 500 Hz, 0.75 g
Survival random vibration 5 to 500 Hz, 2.09 g rms
Functional shock (half-sine, 30 g, 11 ms) Meets the requirements of MIL-PRF-28800F for class
and bench drop test 3 equipment.
EMC Meets the conducted and radiated interference
and immunity requirements of IEC/EN 61326-1.
Meets radiated emission requirements of
CISPR Pub 11/1997 Group 1 class A.
Storage registers Memory is shared by instrument states, user data files,
sweep list files, and waveform sequences. Depending
on the number and size of these files, up to 800 storage
registers and 10 register sequences are available.
Security Display blanking.
Compatibility Agilent Technologies 83550 Series millimeter heads,
Agilent Technologies 8757D Scalar Network Analyzers,
Agilent Technologies EPM Series Power Meters.
Self-test Internal diagnostic routine tests most modules
(including microcircuits) in a preset condition. For each
module, if its node voltages are within acceptable limits,
then the module “passes” the test.
Weight < 22 kg (48 lb.) net, < 30 kg (68 lb.) shipping.
Dimensions 178 mm H x 426 mm W x 498 mm D
(7” H x 16.8” W x 19.6” D in.).
Recommended calibration cycle 24 months
Remote programming
General specifications
1. Storage below –20 °C instrument states may be lost.