6
Level accuracy with step attenuator
1
(dB)
Frequency
> +10 dBm +10 to –10 dBm –10 to –70 dBm –70 to –90 dBm –90 to –110 dBm
250 kHz to 2 GHz
±0.6 ±0.6 ±0.7 ±0.8 ±1.4
> 2 to 20 GHz ±0.8 ±0.8 ±0.9 ±1.0 ±1.7
> 20 to 40 GHz ±1.0 ±0.9 ±1.0 ±2.0
Resolution 0.01 dB
Temperature stability 0.01 dB/°C (typical)
User flatness correction
Number of points 2 to 1601 points/table
Number of tables Up to 10,000, memory limited
Path loss Arbitrary, within attenuator range
Entry modes Remote power meter
2
, remote bus, manual
(user edit/view)
Output impedance 50 Ω (nominal)
SWR (internally leveled) (typical)
250 kHz to 2 GHz < 1.4:1
> 2 GHz to 20 GHz < 1.6:1
> 20 GHz to 40 GHz < 1.8:1
Leveling modes Internal leveling, external detector leveling, millimeter
source module, ALC Off
External detector leveling
Range –0.2 mV to –0.5 V (nominal) (–36 dBm to +4 dBm using
Agilent 33330D/E detector)
Bandwidth 10 kHz (typical) (Note: not intended for pulsed operation)
Maximum reverse power 1/2 Watt (nominal)
1. Specifications apply in CW and List/Step sweep modes over the 15 to 35º C temperature range, with attenuator
hold off (normal operating mode). Degradation outside this range, for ALC power levels > –10 dBm, is typically
< 0.3 dB. In Ramp sweep mode (with Option 007), specifications are typical. For instruments with type-N
connectors (Option 1ED), specifications are degraded typically 0.2 dB above 18 GHz. Level accuracy is not
specified below –110 dBm.
2. Compatible with Agilent Technologies EPM Series (E4418B and E4419B) power meters.