Agilent N5161A/62A/81A/82A/83A MXG Signal Generators Firmware Upgrade Guide
Step 2. Run the Signal Generator’s Functionality Check
5
Step 2. Run the Signal Generator’s Functionality Check
Use the following procedure to confirm that the signal generator powers up and that the internal check identifies no
errors. The internal check evaluates the operation of the signal generator and returns an error message if it detects a
problem.
NOTE The N5161A/62A MXG ATE is identical to the N5181A/82A with the exception that they do not have front
panel functionality (no display or keys). Instead all functionality is controlled via SCPI commands or through
the Web- Enabled MXG. For signal generator functionality, refer to the Users Guide. For information on
Web–Enabled MXG, refer to the Installation Guide, the Programming Guide and the SCPI Command
Reference.
1. Turn on power to the signal generator and let it warm up for at least five minutes.
2. Check to see if the ERR annunciator is on.
•If the ERR annunciator is off, the signal generator’s functionality check passed.
•If the ERR annunciator is on:
a. View the error queue:
Press
Error.
Error message in the error queue appear in the display text area. Refer to the signal generator error
messages list on the documentation CD for information about each error message.
b. Resolve all problems causing errors.
If you are unable to resolve the errors, contact Agilent Technologies (see page 10).
c. Clear the error queue: Press
Clear Error Queue(s) and go to Step 3.
3. Repeat the functionality check:
Cycle the signal generator’s power.
4. Repeat Step 2.