Agilent N5161A/62A/81A/82A/83A MXG Signal Generators Firmware Upgrade Guide
Step 4. Download and Install the Firmware Files
7
Step 4. Download and Install the Firmware Files
From the Agilent Technologies website
1. In an Internet browser, enter http://www.agilent.com/find/upgradeassistant.
2. Download and run the latest Agilent MXG Firmware Package.
3. Once the upgrade assistant starts, follow the program’s prompts (see page 8).
4. After the upgrade completes, press the signal generator’s
Local hardkey to return front panel control.
5. Run the signal generator’s functionality check (see page 5).
From CD–ROM
1. Insert the firmware upgrade CD–ROM into the computer’s CD–ROM drive.
If the MXG FW_Upgrade Assistant fails to auto–execute go to Step 2.
2. Browse the CD for the latest Agilent MXG Firmware Package.
This file will be in the following format example: MXG_FWUpgrade_A0143_Release_200805221340.exe.
3. Double–click the .exe file.
4. Once the upgrade assistant starts, follow the program’s prompts (see page 8).
5. After the upgrade completes, press the signal generator’s
Local hardkey to return front panel control.
6. Run the signal generator’s functionality check (see page 5).
NOTE If the upgrade is interrupted, the instrument tries to load a recovery image; if errors appear, they can be
safely ignored. Run the Upgrade Assistant again to install the proper firmware; additional errors may appear,
but these can also be ignored unless pressing Error > Clear Error Queue does not clear them. If you cannot clear
the errors, contact Agilent Technologies (see page 10).