OPERATING MANUAL
MULTIPLE OUTPUT LINEAR SYSTEM
DC POWER SUPPLY
AGILENT MODELS 6625A, 6626A,
6628A, and 6629A
Agilent Part No 06626-90001
Agilent Model 6625A, Serials 3210A-00664 and Above*
Agilent Model 6626A, Serials 3209A-00824 and Above*
Agilent Model 6628A, Serials 3213A-00233 and Above*
Agilent Model 6629A, Serials 3210A-06721 and Above*
* For instruments with higher Serial Numbers, a change page may be included.
* For instruments with lower Serial Numbers, refer to Appendix E.
Microfiche Part No. 06626-90002 Printed in USA: January, 1993
Updated: February, 2000
Vedere la pagina 0
1 2 3 4 5 6 ... 122 123

Sommario

Pagina 1 - OPERATING MANUAL

OPERATING MANUALMULTIPLE OUTPUT LINEAR SYSTEMDC POWER SUPPLYAGILENT MODELS 6625A, 6626A,6628A, and 6629AAgilent Part No 06626-90001Agilent Model 6625A

Pagina 3 - SAFETY SUMMARY

Alignment Procedures100Alignment Program (continued)2610 Irlo=FNDci(I_range) ! Read low output current2620 !2630 OUTPUT @Ps;"IRHI";

Pagina 4

Alignment Procedures101A

Pagina 5 - DECLARATION OF CONFORMITY

Alignment Procedures102Alignment Program (continued)3650 !3660 !3670 Check_error:SUB Check_error ! Subprogram to check for errors3680 COM /Ps/

Pagina 6 - WHAT THIS MANUAL CONTAINS

Programming with a Series 200/300 Computer103BProgramming With a

Pagina 7 - Table Of Contents

Programming with a Series 200/300 Computer104VOLTAGE AND CURRENT PROGRAMMING WITH VARIABLESYou can use variables in a program to represent dat

Pagina 8 - Table Of Contents (continued)

Programming with a Series 200/300 Computer10510 ASSIGN @Ps TO 705

Pagina 9

Programming with a Series 200/300 Computer106Service Request and Serial PollThe fault and mask registers, when used in conjunction with the se

Pagina 10

Programming with a Series 200/300 Computer107Error DetectionThe p

Pagina 11 - General Information

Programming with a Series 200/300 Computer108LINE 10: Assigns the I/O path name to the power supply.LINE 20: Declares a common block for the

Pagina 12 - Description

Command Summary109CCom

Pagina 13

General Information111General InformationIntroductionThis chapter contains a general description of your power supply as well as its performance speci

Pagina 14 - Basic Operation

Command Summary110Table C-1 Quick Reference Listing of CommandsGeneral Commands************ASTS? < ch >CLRCMODE < 1 (on) or 0 (off) &

Pagina 15 - Specifications

Command Summary111Tabl

Pagina 16

Command Summary112Table C-2 Command Summary (continued)ID? Queries the identification (model number) of the supply. (see page 82).IDATA < c

Pagina 17

Command Summary113Tabl

Pagina 18

Command Summary114Table C-2 Command Summary (continued)ROM? Queries the revision date of the power supply’s firmware. See service manual.SROM?

Pagina 19

Command Summary115Tabl

Pagina 21

Error Codes and Messages117DError

Pagina 22 - General Information

Error Codes and Messages118Table D-2. ERROR Responses (continued)Error Code(ERR? query)Message(ERR key)Explanation3 INVALID STR You sent a com

Pagina 23

Error Codes and Messages119Table D

Pagina 24

General Information12Accessories10833A GP-IB cable, 1 m (3.3 ft)10833B GP-IB cable, 2 m (6.6 ft)10833C GP-IB cable, 4 m (13.2 ft)10833D GP-IB ca

Pagina 26 - Line Fuse

Manual Backdating121

Pagina 27

Agilent Sales and Support Office122Agilent Sales and Support OfficeFor more information about Agilent Technologies test and measurement produc

Pagina 28 - Line Voltage Conversion

123 Manual UpdatesThe following updates have been made to this manual since the print revision indicated on the title page.2/01/00All references to HP

Pagina 29 - GP-IB Interface Connector

General Information13The power supplies allow up to a 10 volt sense lead drop. This feature makes them ideal for test system applications whereremote

Pagina 30

General Information14Basic OperationFigure 1-2 is a block diagram that illustrates the major assemblies contained within the power supply. As sh

Pagina 31 - Getting Started

General Information15Output BoardsThe output boards are linear dc power supplies. Each isolated output operating boundary curve is shown in Figure 1-1

Pagina 32 - 45.153 14.235m

General Information16Output response time: Beginning at the time the power supply has finished processing a VSET command (changeoutput voltage),

Pagina 33

General Information17Table 1-2. Specifications (continued)Source Effect:Voltage 0.5 mV 0.5 mV 0.5 mV 0.5 mV+ Current 0.005 mA 0.005 mA 0.01 mA 0.01 mA

Pagina 34

General Information18Table 1-3. Supplemental Characteristics (continued)Output Power 25 Watt Output 50 Watt OutputOutput Range Lo Range Hi Range

Pagina 35

General Information19Table 1-3. Supplemental Characteristics (continued)High Line Inrush Current: 100 V Opt 120 V Opt 220 V Opt 240 V OptPeak Value 85

Pagina 36 - Turning On Your Supply

2CERTIFICATIONAgilent Technologies certifies that this product met its published specifications at time of shipment from the factory. AgilentTechnolog

Pagina 37

General Information20Table 1-3. Supplemental Characteristics (continued)Weight: Agilent 6625A, 6628A Agilent 6626A, 6629A Net

Pagina 38

General Information21Figure 1-4. CV Operation with Capacitive Load, Stability Graph for all Outputs

Pagina 39

General Information22Figure 1-5. Output Noise (Typical) CC ModeFigure 1-6. Output Noise (Typical) CV Mode

Pagina 40

General Information23Figure 1-7. Output Impedance (Typical) Graph (See Supplemental Characteristics)

Pagina 42

Installation252InstallationIntroductionThis chapter contains instructions for checking and mounting your power supply, connecting your supply to ac po

Pagina 43

Installation26Figure 2-1. Outline DiagramInput Power RequirementsYou can operate this power supply from a nominal 100 V, 120 V, 220 V or 240 V s

Pagina 44

Installation27Figure 2-2. Rear Panel Detail (Agilent 6626A Shown)Table 2-2 Line FusesLineVoltageFuseNeededAgilent Part Number (for ¼ X 1¼ in. fusesonl

Pagina 45 - Output Ranges

Installation28Power CordThe power supply is shipped from the factory with a power cord that has a plug appropriate for your location. Figure 2-4

Pagina 46

Installation29FIRE HAZARD. Make sure the replacement fuse is one of the same type (size) and rating (amps) thatis consistent with the voltage level yo

Pagina 47

3SAFETY SUMMARYThe following general safety precautions must be observed during all phases of operation, service, and repair of thisinstrument. Failur

Pagina 49 - Connecting The Load

Getting Started313Getting StartedIntroductionThis chapter is intended for the first time user of the supply. It provides four main discussions:• Fron

Pagina 50

Getting Started32 Figure 3-1. Agilent 6626A Front Panel Table 3-1. Controls and Indicators Number Controls/Indicators Description Page 1

Pagina 51 - Remote Voltage Sensing

Getting Started33 Table 3-1. Controls and Indicators (continued) Number Controls/Indicators Description Page 4 (cont) UNR- Indicates that the se

Pagina 52

Getting Started34 Table 3-1. Controls and Indicators (continued) Number Controls/Indicators Description Page 7 Output Control Keys(These tw

Pagina 53 - Current Limit Setting

Getting Started35 Table 3-1. Controls and Indicators (continued) Number Controls/Indicators Description Page 7 (cont) Ñ CURRENT - Increases the se

Pagina 54

Getting Started36 Turning On Your Supply The following paragraphs describe the power-on sequence which includes a self test of most of the powe

Pagina 55

Getting Started37 Figure 3-4. Typical Display at Power-On Self-Test Errors If the supply fails the power-on self-test, all power supply outputs wi

Pagina 56 - Parallel Operation

Getting Started38 Voltage Test 1. Set the voltage of the selected output to 10 V by pressing: VSET 1 0 ENTER 2. Check that the display

Pagina 57

Getting Started39 6. Check that the display reads approximately 0 volts and the programmable current limit value. Also, check that the frontpanel CC

Pagina 58

4SAFETY SUMMARY (continued)GENERALAny LEDs used in this product are Class 1 LEDs as per IEC 825-1.ENVIRONMENTAL CONDITIONSThis instrument is intended

Pagina 59 - Series Operation

Getting Started40 Enter/Output Statements The programming statements you use to operate your supply from remote depend on your computer and it

Pagina 60

Getting Started41 Sending a Remote Command To send the power supply a remote command, combine your computer’s output statement with the GP-IB inter

Pagina 61

Getting Started42 Each of these commands is briefly discussed in the following paragraphs to help you get started in programming yoursupply. To

Pagina 62

Getting Started43 Output On/Off. You can turn a specified output on or off. Individual outputs can be controlled as shown below. To turn off output

Pagina 64

Output Connections and Operating Information 45 4 Output Connections and Operating Information Introduction This chapter explains how to make conn

Pagina 65 - Programming Syntax

Output Connections and Operating Information 46 Operating Quadrants Figures 4-2A and 4-2B show the operating locus of your power supply in thr

Pagina 66

Output Connections and Operating Information 47 Figure 4-3 shows the current sink characteristics lower voltages in greater detail. The area in qua

Pagina 67

Output Connections and Operating Information 48 The readback resolution of the 25 watt outputs when metering voltages of 7 volts or below, will

Pagina 68 - Remote Operation

Output Connections and Operating Information 49 UNREGULATED OUTPUT -- the supply informs the user when output regulation is not guaranteed. This can

Pagina 69 - Remote Operation

5DECLARATION OF CONFORMITYaccording to ISO/IEC Guide 22 and EN 45014Manufacturer’s Name: Agilent TechnologiesManufacturer’s Address: 150 Green Pond Ro

Pagina 70

Output Connections and Operating Information 50 Note To prevent tripping of the overvoltage circuit, pick a wire size sufficient to handle the

Pagina 71 - Power Supply Commands

Output Connections and Operating Information 51 Figure 4-4. Optimum Hookup for Multiple Loads, Local Sensing Note When a load is connected through

Pagina 72

Output Connections and Operating Information 52 Figure 4-5. Remote Voltage Sensing The maximum voltage available at the power supply output

Pagina 73

Output Connections and Operating Information 53 Figure 4-6. Total Allowable Load Lead Voltage Drop (total of both leads) with Remote Sensing Outpu

Pagina 74

Output Connections and Operating Information 54 Open Sense Leads The sense leads are part of the supply’s feedback path. Connect them in such

Pagina 75

Output Connections and Operating Information 55 Figure 4-8. External Trigger Circuit Figure 4-9. Equivalent Internal OV Trigger Circuit Power Su

Pagina 76

Output Connections and Operating Information 56 Figure 4-10. Recommended Protection Circuit for Battery Charging Capacitive Load Limitation

Pagina 77

Output Connections and Operating Information 57 Figure 4-11. Parallel Connections with Local Sensing CV Operation For CV operation, one output mus

Pagina 78

Output Connections and Operating Information 58 CC Operation For CC operation, set the output voltages as outlined in CV operation (page 57),

Pagina 79

Output Connections and Operating Information 59 Series Operation SHOCK HAZARD. Floating voltages must not exceed 240 Vdc. No output terminal may be

Pagina 80

6WHAT THIS MANUAL CONTAINSIt contains information relating to the installation, operation, and programming of these supplies as outlined below.Mainten

Pagina 81

Output Connections and Operating Information 60 CC Operation For CC operation, the current setting of each output must be programmed to the de

Pagina 82

Output Connections and Operating Information 61 Voltage All series specifications referring to voltage are twice the single output specificationex

Pagina 84

Remote Operation 63 5 Remote Operation Introduction Chapter 3 introduced you to the basics of remote operation and provided a few simple examples

Pagina 85 - Local Operation

Remote Operation 64the SRQ annunciator regardless of whether the condition that caused the service request continues to exist. The servicereque

Pagina 86

Remote Operation 65 Power-On Service Request (PON) The power supply can request service from the controller when the power is turned on. This reques

Pagina 87

Remote Operation 66 Table 5-1. Power Supply Commands (continued) Command Header OutputChannel Data Range (Fig. 5-2) Syntax Query present hard

Pagina 88

Remote Operation 67 Table 5-1. Power Supply Commands (continued) Command Header OutputChannel Data Range (Fig. 5-2) Syntax Store present output sta

Pagina 89

Remote Operation 68 Figure 5-2 (Sheet 1 of 2). Syntax Forms For Power Supply Commands

Pagina 90

Remote Operation 69 Figure 5-2 (Sheet 2 of 2). Syntax Forms For Power Supply Commands

Pagina 91 - Alignment Procedures

7Table Of Contents1 General InformationIntroduction... ...

Pagina 92

Remote Operation 70 Table 5-2. Power Supply Queries Query Header (Note 7) Channel(Note 1) Response (Notes 5 and 6) Initial Value Syntax (Fig.

Pagina 93

Remote Operation 71 Initial Conditions Immediately after power on from the factory, the power supply automatically undergoes a self-test and sets al

Pagina 94 - Figure A-5

Remote Operation 72 VSET 1,.45 If the output channel is operating in constant voltage mode (CV annunciator on) then the actual voltage is the

Pagina 95

Remote Operation 73 Current Programming To program the current, send the output channel and the programmed value in amps. In the example below, outp

Pagina 96

Remote Operation 74 Output On/Off The OUT command disables/enables an output channel of the power supply. It will not disturb any other progr

Pagina 97

Remote Operation 75 Overcurrent Protection (OCP) The OCP is a protection feature employed by the power supply to guard against excessive output curr

Pagina 98

Remote Operation 76 The Clear Command This command will return the power supply to its power-on state and all parameters are returned to thei

Pagina 99

Remote Operation 77 Table 5-5. Bit Assignment for the Status, Astatus, Fault, and Mask Registers Bit Position 7 6 5 4 3 2 1 0 Bit Weight 128

Pagina 100

Remote Operation 78 UNMASK 2,XXX where XXX specifies the numeric code (0 to 255) for the unmasked conditions (see Table 5-5). If during operat

Pagina 101

Remote Operation 79 Service Request Generation When operating your supply, you may want it to request service every time a fault or a programming e

Pagina 102

8Table Of Contents (continued) Positive and Negative Voltages...

Pagina 103 - I/O Path Names

Remote Operation 80 The ability to generate service requests can be enabled or disabled using the SRQ command as described below. To disable

Pagina 104 - Voltage and Current Readback

Remote Operation 81Reprogramming delay will delay the onset of certain fault conditions and prevent the power supply from registering a faultwhen the

Pagina 105

Remote Operation 82Other QueriesIn the examples discussed above, you saw how to use queries for each function discussed. The following paragrap

Pagina 106

Remote Operation 83To enable all outputs in constant current mode at turn on send: DCPON 2To disable all outputs in constant current mode at turn-on

Pagina 107

Remote Operation 84Table 5-8. Error Messages (continued)Front PanelResponseGP-IBCodeExplanation HDW ERR CH 1 11 Errors 11 through 14 refer to a

Pagina 108

Local Operation856Local OperationIntroductionChapter 3 introduced you to the supply’s front panel controls and indicators to help you turn on the supp

Pagina 109 - Command Summary

Local Operation86Figure 6-1. Front Panel (Model 6626A shown)Pressing the ENTER key will enter the values displayed for the function indicated, i

Pagina 110

Local Operation87The voltage step keys can also be used to change the voltage setting.VSET ÑVOLT or ÒVOLT then ENTERThe voltage can be set in the imme

Pagina 111

Local Operation88Setting Overvoltage ProtectionProgrammable overvoltage protection (OVP) guards your load against overvoltage by crowbarring and

Pagina 112

Local Operation89For example, you can change the address of your supply to 10 by pressing:ADDR 1 0 ENTERDisplaying Error MessagesThe power supply can

Pagina 113

9Table Of Contents (continued)Local Control Of Output Functions...

Pagina 115

Alignment Procedures91AA

Pagina 116

Alignment Procedures92The program contains embedded comments (identified by a leading asterisk {! } ) which explain various sections andproced

Pagina 118

Alignment Procedures94Figure A-3Figure A-4Figure A-5

Pagina 119

Alignment Procedures95Al

Pagina 120

Alignment Procedures96Alignment Program (continued)530 ! -1 is reverse polarity, this saves540 ! reversing leads as sink channel550 ! becomes

Pagina 121 - Manual Backdating

Alignment Procedures97Al

Pagina 122

Alignment Procedures98Alignment Program (continued)1570 PRINT "Press CONTINUE when ready1580 PAUSE1590 Output_on1600 !1610 Cal_v: !1

Pagina 123 - Manual Updates

Alignment Procedures99Al