Module 13
SPGU Control and Applications
13-16
3-Level Pulses
Every channel can output 3-level pulse
80 Vp-p within +/-40 V window (at open load)
Pulse width and level of each channel can be set individually.
Pulse Generator Mode
+40V
80Vpp
-40V
SPGU10
SPGU1
SPGU3
SPGU2
Up to 10 output channels can be installed in one B1500A and each channel can output 3-level pulse with a
maximum output voltage.
The channels are independent. So the channels can apply the different pulses. However, the pulse period
value is common for all channels.