Using the Oscilloscope
Calibrating the oscilloscope
4 Calibrate the Logic Trigger of the Self Cal.
a Obtain a BNC 50-W, 48-inch cable.
b Select Start. The instrument will prompt you to connect the
cable to the appropriate location on the rear panel of the
instrument.
c Select the Procedure field, then select Logic Trigger from
the pop-up menu.
d Select the Start field and follow the instructions on the
display.
e After completion of the Logic Trigger calibration, remove the
cable from the instrument.
Protect the operational accuracy calibration
factors
Facing the front panel, reach around to the lower right corner on the back
of the logic analyzer. Flip the PROTECT/UNPROTECT switch to
PROTECT (flip the switch down).