Agilent Technologies DP235 Guida Utente Pagina 556

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
Vedere la pagina 555
556
Troubleshooting the LAN Connection
Solutions to Common Problems
If you cannot copy files from the logic analyzer
If you can only copy a few bytes of a file:
Copy the file of interest to your PC or workstation, and use the new, local
copy as your working copy. Some applications cannot work directly with
the dynamic files from the logic analyzer.
If you are on a UNIX workstation, try using dd instead of cp.
If you cannot restore raw files
If the setup.raw and data.raw files seem to be ignored when you copy
them to the logic analyzer:
Verify the LAN connection between your computer and the logic analyzer.
Refer to "If you cannot connect to the logic analyzer" earlier in this section.
Change the file time-out. To change the time-out:
1 Go to the logic analyzer System External I/O menu.
2 Select the LAN Settings field.
3 Set the file time out.
A value of about 1.5 seconds should be sufficient. Set the time out to
the minimum value that will allow network operations to work reliably.
Vedere la pagina 555
1 2 ... 551 552 553 554 555 556 557 558 559 560 561 ... 627 628

Commenti su questo manuale

Nessun commento