Agilent Technologies 22A Specifiche Pagina 76

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 120
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 75
76
The final factor in dynamic range is the phase noise on our spectrum analyzer
LO, and this affects only third-order distortion measurements. For example,
suppose we are making a two-tone, third-order distortion measurement on
an amplifier, and our test tones are separated by 10 kHz. The third-order
distortion components will also be separated from the test tones by 10 kHz.
For this measurement we might find ourselves using a 1 kHz resolution
bandwidth. Referring to Figure 6-3 and allowing for a 10 dB decrease in
the noise curve, we would find a maximum dynamic range of about 88 dB.
Suppose however, that our phase noise at a 10 kHz offset is only –80 dBc.
Then 80 dB becomes the ultimate limit of dynamic range for this measurement,
as shown in Figure 6-4.
In summary, the dynamic range of a spectrum analyzer is limited by three
factors: the distortion performance of the input mixer, the broadband noise
floor (sensitivity) of the system, and the phase noise of the local oscillator.
60 50 40 30 20 10 +100
100
110
90
80
70
60
50
40
30
20
10
Mixer level (dBm)
(dBc)
Dynamic range
reduction due
to phase noise
Phase noise
(10 kHz offset)
Figure 6-4. Phase noise can limit third-order intermodulation tests
Vedere la pagina 75
1 2 ... 71 72 73 74 75 76 77 78 79 80 81 ... 119 120

Commenti su questo manuale

Nessun commento