Characteristics of the
1146A Current Probe
Bandwidth* dc to 100 kHz (-3 dB)
Current Range* 100 mV/A:100 mA to
10 A peak
10 mV/A:1 to 100 A peak
Output Signal 1000 mV peak max
AC Current Accuracy*
Range: 100 mV/A (50 mA to
10 A peak)
Accuracy: 3% of reading ±50 mA
Range: 10 mV/A (500 mA to
40 A peak)
Accuracy: 4% of reading ±50 mA
Range: 10 mV/A (40 A to
100 A peak)
Accuracy: 15% max at 100 A
Phase Shift < 1° from dc to 65 Hz on
10 mV/A
< 1.5° from dc to 65 Hz on
100 mV/A
Noise Range 10 mV/A: 480 µV
Range 100 mV/A: 3 mV
Slew Rate Range 10 mV/A: 20 mV/µs
Range 100 mV/A: 0.3 V/µs
Insertion Impedance (50/60 Hz) 0.01
Rise or Fall Time Range 100 mV/A: 3 µs
Range 10 mV/A: 4 µs
Maximum 600 Vrms max.
Working Voltage
Maximum Common 600 Vrms max.
Mode Voltage
Influence of <0.2 mA/A AC
Adjacent Conductor
Influence of 0.5% of reading at
Conductor Position 1 kHz in jaw
Battery 9 V alkaline (NEDA 1604A,
IEC 6LR61)
Low battery green LED when >6.5 V
Battery Life 55 hours typical
* Characteristics marked with asterisks are specified
performance. Others are typical characteristics.
Note: Reference conditions 23°C ± 5°C, 20 to 75% rela-
tive humidity, dc to 1 kHz, probe zeroed, 1-minute
warmup, batteries at 9 V + 0.1 V, external magnetic field
<40 A/m, no dc component, no external current carrying
conductor, 1 M/ 100 pF load, conductor centered in jaw.
Characteristics of
N2774A Current Probe
Bandwidth (-3 dB) dc to 50 MHz
Risetime 7 ns or less
Rated Current 15 A peak (ac+dc
components)
Maximum Peak 30 A peak;
Current Non-continuous 50 A peak;
at pulse width at 10 µs
Output Voltage Rate 0.1 V/A
Amplitude ±0.5% rdg, ±1 mV (dc and
Accuracy 45 to 66 Hz, rated current)
Noise Equivalent to 2.5 mArms or
less (for 20 MHz bandwidth
measuring instrument)
Temperature ±2% or less (within a
Coefficient range of 0 °C to 40 °C or
for Sensitivity 32 °F to 104 °F)
Effect of External Equivalent to a maximum
Magnetic Fields of 20 mA (in a dc to 60 Hz,
400 A/m magnetic field)
Maximum 3 VA (with rated current)
Rated Power
Rated Supply Voltage dc ±12 V ±1 V
Diameter of 5 mm dia. (0.2 dia.)
Measurable
Conductors
Cable Lengths Sensor cable:
Appox. 1.5 m (59.0)
Power supply cable:
Appox. 1 m (39.4)
Note: The above specifications are guaranteed at
23 °C ± 3 °C (or 73 °F ± 5 °F)
Current Probes (continued)
Ordering information for Agilent
1146A and N2774A current probe
1146A 100 kHz current probe
N2774A 50 MHz current probe
N2775A Power supply for N2774A