Characteristics of the
1146A Current Probe
Bandwidth* dc to 100 kHz (–3 dB)
Current Range* 100 mV/A:100 mA to
10 A peak
10 mV/A:1 to 100 A peak
Output Signal 1000 mV peak max
AC Current Accuracy*
Range: 100 mV/A (50 mA to
10 A peak)
Accuracy: 3% of reading ±50 mA
Range: 10 mV/A (500 mA to
40 A peak)
Accuracy: 4% of reading ±50 mA
Range: 10 mV/A (40 A to
100 A peak)
Accuracy: 15% max at 100 A
Phase Shift < 1° from dc to 65 Hz on
10 mV/A
< 1.5° from dc to 65 Hz on
100 mV/A
Noise Range 10 mV/A: 480 µV
Range 100 mV/A: 3 mV
Slew Rate Range 10 mV/A: 20 mV/µs
Range 100 mV/A: 0.3 V/µs
Insertion Impedance (50/60 Hz) 0.01
Rise or Fall Time Range 100 mV/A: 3 µs
Range 10 mV/A: 4 µs
Maximum 600 Vrms max.
Working Voltage
Maximum Common 600 Vrms max.
Mode Voltage
Influence of < 0.2 mA/A AC
Adjacent Conductor
Influence of 0.5% of reading at
Conductor Position 1 kHz in jaw
Battery 9 V alkaline (NEDA 1604A,
IEC 6LR61)
Low battery green LED when > 6.5 V
Battery Life 55 hours typical
* Characteristics marked with asterisks are specified
performance. Others are typical characteristics.
Note: Reference conditions 23°C ± 5°C, 20 to 75% rela-
tive humidity, dc to 1 kHz, probe zeroed, 1-minute
warmup, batteries at 9 V + 0.1 V, external magnetic field
<40 A/m, no dc component, no external current carrying
conductor, 1 M/ 100 pF load, conductor centered in jaw.
Characteristics of
1147A/N2774A Current Probe
Bandwidth (-3 dB) dc to 50 MHz
Risetime 7 ns or less
Rated Current 15 A peak (ac+dc
components)
Maximum Peak 30 A peak;
Current Non-continuous
50 A peak;
at pulse width of
10 µs or less
Output Voltage Rate 0.1 V/A
Amplitude ±1% rdg, ±1 mV (dc and
Accuracy 45 to 66 Hz, rated current)
Noise Equivalent to 2.5 mArms or
less (for 20 MHz bandwidth
measuring instrument)
Temperature ±2% or less (within a
Coefficient range of 0 °C to 40 °C or
for Sensitivity 32 °F to 104 °F)
Effect of External Equivalent to a maximum
Magnetic Fields of 20 mA (in a dc to 60 Hz,
400 A/m magnetic field)
Maximum 3 VA (with rated current)
Rated Power
Rated Supply Voltage dc ±12 V ±1 V
Diameter of 5 mm dia. (0.2” dia.)
Measurable
Conductors
Probe Interface BNC (N2774A)
AutoProbe interface
(1147A)
Cable Lengths Sensor cable:
Appox. 1.5 m (59.0”)
Power supply cable:
Appox. 1 m (39.4”)
Maximum Number 2 (1147A)
of Probes Supported 4 (N2774A with 2 N2775A)
Note: The above specifications are guaranteed at
23 °C ± 3 °C (or 73 °F ± 5 °F)
Current Probes (continued)
Ordering information for Agilent
current probes
1146A 100 kHz current probe
N2774A 50 MHz current probe
N2775A Power supply for N2774A
1147A 50 MHz current probe
with AutoProbe interface