3
Agilent 35670A
Specifications
Instrument specifications apply after
15 minutes warm-up and within 2 hours of the
last self-calibration. When the internal cooling
fan has been turned OFF, specifications apply
within 5 minutes of the last self-calibration.
All specifications are with 400 line frequency
resolution and with anti-alias filters enabled
unless stated otherwise.
Frequency
Maximum Range**
1 Channel Mode 102.4kHz,
51.2 kHz (option AY6*)
2 Channel Mode 51.2 kHz
4 Channel Mode (option AY6 only) 25.6 kHz
Spans
1 Channel Mode 195.3 mHz to 102.4 kHz
2 Channel Mode 97.7 mHz to 51.2 kHz
4 Channel Mode (option AY6 only) 97.7 mHz to 25.6 kHz
Minimimum Resolution
1 Channel Mode 122 µHz (1600 line display)
2 Channel Mode 61 µHz (1600 line display)
4 Channel Mode (option AY6 only) 122 µHz (800 line display)
Maximum Real-Time Bandwidth
(FFT Span for Continuous Data Acquistion) (Preset, Fast Averaging)
1 Channel Mode 25.6 kHz
2 Channel Mode 12.8 kHz
4 Channel Mode (option AY6 only) 6.4 kHz
Measurement Rate
(Typical) (Preset, Fast Averaging)
1 Channel Mode ≥ 70 Averages/Second
2 Channel Mode ≥ 33 Averages/Second
4 Channel Mode (option AY6 only) ≥ 15 Averages/Second
Display Update Rate
Typical
(Preset, Fast Average OFF)
≥ 5 Updates/Second
Maximum ≥ 9 Updates/Second
(Preset, Fast Average Off, Single Channel,
Single Display, Undisplayed Trace Displays
set to Data Registers)
Accuracy ±30 ppm (.003%)
* Option AY6 single channel maximum range extends
to 102.4 kHz without anti-alias filter protection.
** Show All Lines mode allows display of up to 131.1,
65.5 and 32.7 kHz respectively. Amplitudes accura-
cy is unspecified and not alias protected.
Single Channel Ampltude
Absolute Amplitude Accuracy (FFT)
(A combination of Full Scale Accuracy, Full
Scale Flatness, and Amplitude Linearity.)
±2.92% (0.25dB) of Reading
±0.025% of Full Scale
FFT Full Scale Accuracy at 1 kHz (0 dBfs)
±0.15 dB (1.74%)
FFT Full Scale Flatness (0 dBfs)
Relative to 1 kHz
±0.2 dB (2.33%)
FFT Amplitude Linearity at 1 kHz
Measured on +27 dBVrms range with time avg,
0 to -80 dBfs
±0.58% (0.05dB) of reading ±0.025%
of full scale
Amplitude Resolution
(16 bits less 2 dB over-range) with averaging
0.0019% of full scale (typical)
Residual DC Response (FFT Mode)
Frequency Display (Excludes A-weight filter)
<-30 dBfs or 0.5 mVdc