Agilent Technologies 35670A Manuale Utente Pagina 7

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 8
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 6
7
Computed Order Tracking - Option 1D0
Maximum Order x Maximum RPM
60
Online (Real Time) 1 Channel Mode 25,600 Hz
2 Channel Mode 12,800 Hz
4 Channel Mode 6,400 Hz
Capture Playback 1 Channel Mode 102,400 Hz
2 Channel Mode 51,200 Hz
4 Channel Mode 25,600 Hz
Number of Orders
200 5 RPM 491,519
(Maximum useable RPM is limited by
Resolution, Tach Pulse Rate,Pulses/Revolution
and Average Mode Settings.)
Delta Order 1/128 to 1/1
Resolution 400
(Maximum Order) / (Delta Order)
Maximum RPM Ramp Rate 1000 RPM / second real-time (typical)
1000 - 10,000 RPM Run Up
Maximum Order 10
Delta Order 0.1
RPM Step 30 (1 Channel)
60 (2 Channel)
120 (4 Channel)
Order Track Amplitude Accuracy ±1 dB (typical)
Real Time Octave Analysis - Option 1D1
Standards Conforms to ANSI Standard S1.11 - 1986,
Order 3, Type 1-D, Extended and Optional
Frequency Ranges
Conforms to IEC 651-1979 Type 0 Impulse ,
and ANSI S1.4
Frequency Ranges
(at centers)
Online (Real Time):
Single Channel 2 Channel 4 Channel
1/1 Octave 0.063 - 16 kHz 0.063 - 8 kHz 0.063 - 4 kHz
1/3 Octave 0.08 - 40 kHz 0.08 - 20 kHz 0.08 - 10 kHz
1/12 Octave 0.0997 - 12.338 kHz 0.0997 - 6.169 kHz 0.0997 - 3.084 kHz
Capture Playback:
1/1 Octave 0.063 - 16 kHz 0.063 - 16 kHz 0.063 - 16 kHz
1/3 Octave 0.08 - 31.5 kHz 0.08 - 31.5 kHz 0.08 - 31.5 kHz
1/12 Octave 0.0997 - 49.35 kHz 0.0997 - 49.35 kHz 0.0997 - 49.35 kHz
One to 12 octaves can be measured and displayed.
1/1-, 1/3-, and 1/12-octave true center frequencies related by the formula: f(i+1)/f(i) = 2^(1/n); n=1,
3, or 12; Where 1000 Hz is the reference for 1/1, 1/3 Octave, and 1000*2^(1/24) Hz is the reference
for 1/12 octave. The marker returns the ANSI standard preferred frequencies.
Accuracy
1 Second Stable Average
Single Tone at Band Center: ± 0.20 dB
Readings are taken from the Linear Total Power Spectrum Bin.
It is derived from sum of each filter.
1/3-Octave Dynamic Range > 80 dB (typical) per ANSI S1.11-1986
2 Second Stable Average
Total power limited by input noise level
(
)
Swept Sine Measurements - Option 1D2
Dynamic Range 130 dB
Tested with 11 dBVrms
source level at: 100 mSec
integration
Arbitrary Waveform Source - Option 1D4
Amplitude Range AC: ±5V peak*
DC: ±10V*
* Vac
pk
+ |Vdc| 10V
Record Length # of Points = 2.56 x
Lines of Resolution,
or # of Complex
Points = 1.28 x Lines
of Resolution
DAC Resolution
0.2828 Vpk to 5 Vpk 2.5 mV
0 Vpk to 0.2828 Vpk 0.25 mV
Vedere la pagina 6
1 2 3 4 5 6 7 8

Commenti su questo manuale

Nessun commento