Performance Tests Performance Test Instructions
86 Agilent 81480A and 81680A, 40A, 82A, 42A, & 89A Tunable Laser Modules Users Guide,
Sixth Edition
f For l
FBG
± 1 nm, check and note t he power level displayed by the
OSA at every 0.1 nm interval. That is, fill out the table shown in
Table Table 19.
3 Determine minimum value of filter transmission and actual Fiber-
Bragg-Grating center wavelength, λ
0
.
a Check for minimum transmitted peak power in Table Table 19.
b Mark the associated wavelength set on the Tunable Laser,
TLS_l
0
, and note the value in the test record.
Table 19 Filter Transmission Characteristic
Tunable Laser Module Output
Wavelength Relative to λ
FBG
Peak Power Level
Associated Wavelength Dis-
played on OSA
1.0 nm dBm nm
0.9 nm dBm nm
0.8 nm dBm nm
0.7 nm dBm nm
0.6 nm dBm nm
0.5 nm dBm nm
0.4 nm dBm nm
0.3 nm dBm nm
0.2 nm dBm nm
0.1 nm dBm nm
±0nm = λ
FBG
dBm nm
+0.1 nm dBm nm
+0.2 nm dBm nm
+0.3 nm dBm nm
+0.4 nm dBm nm
+0.5 nm dBm nm
+0.6 nm dBm nm
+0.7 nm dBm nm
+0.8 nm dBm nm
+0.9 nm µW nm
+1.0 nm µW nm