Chapter 2 159
Making Measurements
Example 13: Making Pulsed RF Measurements
6. Increase the sweep time (that is, the sweep becomes slower) until the
display fills in and becomes a solid line (see Figure 2-84). If this line
does not fill in, the instrument is not in broadband mode, in which
case the following procedures for side lobe ratio, pulse width, and
peak pulse power do not apply. For further reference, consult Agilent
Technologies Application Note 150-2, entitled "Pulsed RF"
(literature number 5954-2705).
Figure 2-83 Main Lobe and Side Lobes
Vedere la pagina 158