Probes and Accessories
10070C 10074C 10076A N2771A
(shipped with scope) high-voltage probe high-voltage probe
Probe Bandwidth 20 MHz 150 MHz 250 MHz 50 MHz
Probe Risetime (Calculated) < 17.5 ns < 2.33 ns < 1.4 ns < 7 ns
Attenuation Ratio 1:1 10:1 100:1 1000:1
Input Resistance
(when terminated into 1 M) 1 M 10 M 66.7 M 100 M
Input Capacitance Approx 70 pF Approx 15 pF Approx. 3 pF Approx. 1 pF
Maximum Input (dc + peak ac) 400 Vpk CAT I 500 Vpk CAT I 4000 Vpk 15 kV dc, 10 kVrms
(mains isolated) (mains isolated) 30 kV dc + peak ac
400 Vpk CAT II 400 Vpk CAT II
(post receptacle mains) (post receptacle mains)
Compensation Range None 9 – 17 pF 6 – 20 pF 7 – 25 pF
Probe readout Yes Yes Yes no
To get the most out of your
scope, you need the right
probes and accessories for
your particular application.
That's why Agilent Technologies
offers a complete family of
innovative probes and
accessories for 54600-Series
scopes. For more
comprehensive information,
please see the Agilent
54600-Series Oscilloscopes
Probes and Accessories
data sheet or visiting our
website at
www.agilent.com/find/megazoom
.
Passive Probes
Agilent 10070-family passive probes are
high-quality, general-purpose probes
designed for optimal performance with
54600-Series oscilloscopes. Ruggedized
for general-purpose measurements,
these probes are tested to ensure that
they operate in the toughest of
conditions. They come with a variety of
accessories to make your job easier.
Mixed Signal Oscilloscope Logic Probes
The probes for the 54621D and 54622D
MSOs are the same ones used with
Agilent industry-leading high-
performance logic analyzers. This
means we can offer the best
performance, great value and access to
the industry's broadest range of logic
probing accessories.
The 10089A 2 x 8-signal logic probe
with flying leads makes it possible to
connect at several different places on
your DUT. The probe cable provides
two sets of eight channels so you can
work conveniently with only one set if
that's all you require. This probe is
included with your MSO scope.
The 10085A 16-channel logic probe
and termination adapter is designed
to make it easy to connect to industry-
standard 20-pin board connectors.
Agilent Passive Probes Selection Guide