Strobe output and trigger output
Strobe: Marks the beginning of each
burst
Trigger format: One pulse per period
with 50% duty cycle typical
External mode: 9 ns typical
Level: TTL or ECL selectable
Output impedance: 50 Ω typical
Max. external voltage: -2 V/+7 V
Transition times: 1.0 ns typical for TTL,
600 ps typical for ECL
User Interface
Overprogramming: Values can be
entered, exceeding the specified
range, to fully exploit the hardware
limits.
Setting check: Warning messages
indicate potentially conflicting param-
eters due to inaccuracy. Error mes-
sages indicate conflicting parameters.
Help key: Displays a context sensitive
message.
Autoset key: Resolves all timing
conflicts.
Non-volatile memory: The current
setting is saved on power-down. Up
to nine user settings and one fixed
default setting are stored in the
instrument.
Memory card: 99 settings can be stored
on a 1 MB PCMCIA card (MS-DOS).
Remote Control: Operates according to
IEEE standard 488.2, 1987 and SCPI
1992.0
Function Code: SH1,AH1,T6,L4,SR1,
RL1,PP0,DC1,DT1,C0
Programming times: All checks and
display off
Typical delay times Agilent 81101A
Instrument mode From To Typical value
External width EXT. INPUT STROBE/TRIGGER OUT 8.5 ns
OUTPUT 1/OUTPUT 2 22.5 ns
All other modes EXT. INPUT/CLK INPUT STROBE/TRIGGER OUT 12.0 ns
OUTPUT 1/OUTPUT 2 29 ns
STROBE/TRIGGER OUT OUTPUT 1/OUTPUT 2 17 ns
ASCII command Typical exec. time
One parameter or mode 30 ms typical
Recall setting 250 ms typical
4